全自动酶偶联法测定腺苷脱氨酶 被引量:13

Automated enzyme-coupled assay for determination of adenosine desaminase
在线阅读 下载PDF
导出
摘要 目的 建立全自动测定腺苷脱氨酶 (ADA)的酶偶联分光光度法。方法 用酶联比色法对pH值、腺苷浓度等反应条件及各种实验参数进行研究。用该法测定了 90例健康者的血清样本及 99例胸腹水的ADA活性。结果 该法最低检出限为1.7U/L ,线性范围可达 2 2 0U/L(r=0 .9989) ,批内CV为 2 9%~ 5 .4 % ,批间CV为 4 .7%~ 6 4 % ,回收率为 99.7%。胆红素 <136 .8μmol/L ,血红蛋白 <3.5g/L ,抗坏血酸 <2 84 μmol/L和三酰甘油 <9.0 3mmol/L的情况下 ,对测定结果无显著性影响。 结论 本法灵敏度高 ,线性、精密度好 ,结果准确 ,干扰少 ,且操作简便快速 ,适用于全自动生化分析仪分析。 Objective To establish an automated enzyme-coupled spectrophotometric assay for determination of adenosine desaminase (ADA).Methods The reaction conditions,e.g.,pH,substrate concentration and so on were optimized,and the parameters in the assay were set up.The serum levels of health individual (n=90) and ADA levels in the pleural fluid or ascites (n=99) were determined and the assay was evaluated.Results The calibration curve for adenosine desaminase was linear between 1.7U/L and 220U/L.The inter and intra CV were 5.35% and 6.38% respectively.The mean recovery rate was 99.7%. This method was free from interference by brilirubin (<136.8 μmol/L),hemoglobin (<3.5g/L),ascorbate (<284 μmol/L) and triglycerides (<9.03mmol/L).Conclusion The sensitivity,linearity and precisions of the established assay were satisfied without noticeable interference.The procedure was accurate,rapid and automatable to be used in clinical assay.
出处 《临床检验杂志》 CAS CSCD 北大核心 2004年第5期338-340,共3页 Chinese Journal of Clinical Laboratory Science
关键词 腺苷脱氨酶 酶偶联法 ADA 胸腹水 三酰甘油 胆红素 CV 线性 最低检出限 反应条件 adenosine desaminase enzyme-coupled spectrophotometic assay continous-monitoring method
  • 相关文献

参考文献3

  • 1王坤 瞿韬 左砚芩.实用诊断酶学[M].科学技术文献出版社重庆分社,1988.610-618.
  • 2Guisti G. Adenosine deaminase [ A ]. In: Bergmyer H U. Methods of enzymatic analysis [ M ]. 12nd ed. New York :Acad Press, 1974. 1092-1099.
  • 3陈敬捷,杨小兵,秦长明.腺苷脱氨酶的全自动分析方法的建立[J].临床荟萃,2001,16(13):615-617. 被引量:3

二级参考文献3

  • 1杨振华.腺苷脱氨酶测定方法[J].南通医学院学报,1981,1:82-4.
  • 2[2]Guisti G.Adenosine deaminase.In:Bergmyer IIV et al.Methods of enzymatic analysis.12nd ed.New York:Acad Press,1974.1092-1099
  • 3石寅皓.全自动生化分析仪限额参数的设置[J].临床检验杂志,1998,16(2):95-96. 被引量:16

共引文献5

同被引文献59

引证文献13

二级引证文献48

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部