期刊文献+

Estimation of measuring uncertainty for optical micro-coordinate measuring machine 被引量:1

Estimation of measuring uncertainty for optical micro-coordinate measuring machine
原文传递
导出
摘要 Based on the evaluation principle of the measuring uncertainty of the traditional coordinate measuring machine (CMM), the analysis and evaluation of the measuring uncertainty for optical micro-CMM have been made. Optical micro-CMM is an integrated measuring system with optical, mechanical, and electronic components, which may influence the measuring uncertainty of the optical micro-CMM. If the influence of laser speckle is taken into account, its longitudinal measuring uncertainty is 2.0 μm, otherwise it is 0.88 μm. It is proved that the estimation of the synthetic uncertainty for optical micro-CMM is correct and reliable by measuring the standard reference materials and simulating the influence of the diameter of laser beam. With Heisenberg's uncertainty principle and quantum mechanics theory, a method for improving the measuring accuracy of optical micro-CMM through adding a diaphragm in the receiving terminal of the light path was proposed, and the measuring results are verified by experiments. Based on the evaluation principle of the measuring uncertainty of the traditional coordinate measuring machine (CMM), the analysis and evaluation of the measuring uncertainty for optical micro-CMM have been made. Optical micro-CMM is an integrated measuring system with optical, mechanical, and electronic components, which may influence the measuring uncertainty of the optical micro-CMM. If the influence of laser speckle is taken into account, its longitudinal measuring uncertainty is 2.0 μm, otherwise it is 0.88 μm. It is proved that the estimation of the synthetic uncertainty for optical micro-CMM is correct and reliable by measuring the standard reference materials and simulating the influence of the diameter of laser beam. With Heisenberg's uncertainty principle and quantum mechanics theory, a method for improving the measuring accuracy of optical micro-CMM through adding a diaphragm in the receiving terminal of the light path was proposed, and the measuring results are verified by experiments.
作者 宋康 蒋庄德
出处 《Chinese Optics Letters》 SCIE EI CAS CSCD 2004年第12期704-707,共4页 中国光学快报(英文版)
基金 ThisworkwassupportedbytheNational"973"ProgramofChinaunderGrantNo.2003CD716203.
关键词 Coordinate measuring machines DIAPHRAGMS Laser beams Measurement theory MEASUREMENTS Optical instruments SPECKLE Coordinate measuring machines Diaphragms Laser beams Measurement theory Measurements Optical instruments Speckle
  • 相关文献

同被引文献7

引证文献1

二级引证文献3

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部