超声对高强度聚焦超声治疗的实时监控研究 被引量:7

Study on real-time ultrasonography supervise of high intensity focused ultrasound therapy
原文传递
导出
摘要 目的 探讨在高强度聚焦超声(HIFU)治疗过程中,实时监控超声的图像变化与HIFU辐照产生的凝固性坏死灶大小之间的相关性。方法 观测在相同声强和辐照时间、治疗深度等指标情况下,HIFU定点辐照家兔肝脏,靶区辐照前及治疗结束后即刻、2min、5min、10min后的超声声像图变化和灰度值变化,进而观测超声声像图面积与生物学焦域面积的关系。结果 HIFU辐照后靶区超声回声强度明显增强,灰度值增高,但随着时间延长,灰度则逐渐减弱,声像图面积也随时间逐渐减小,但在 2min时趋于稳定。而损伤的凝固性坏死灶最大切面面积与不同时间的声像图面积比较,介于 2min和 5min的声像图面积之间,与两者比较,差异均无显著性意义(P>0. 05)。结论 通过观察HIFU活体治疗时的灰度值和声像图变化,可间接指导HFIU治疗剂量的选择和判断治疗的效果,为临床HIFU实时监控治疗提供理论和实验基础。 Objective To explore the relativity of the ultrasonic image observed in real time and the size of coagulative necrosis produced by the exposure with high intensity focused ultrasound(HIFU).Methods The rabbit liver tissues were dot-exposured with HIFU with same intensity,duration series and depth,the ultrasonography prior to exposure and the instantaneous,second minute,fifth minute,and tenth minute posterior to exposure were inspected.The gray values were measured,the size of ultrasonic image of different time posterior to exposure and the size of coagulative necrosis area were compared.Results A bright hyperecho spot was produced immediately after exposure with HIFU.And the gray value was increased obviously, then the gray scale weared off gradually and went to a stable level at the second minute.The size of coagulative necrosis area compared with the size of ultrasonic image,was smaller at the second minute and bigger at the fifth minute,but all of these differences were not significant (P>(0.05)).Conclusions The changes of the gray scale and the ultrasonic image area during the HIFU therapy could be used as guidance to choose the HIFU dosage and to judge curative effect of the HIFU.The study provided a theoretic and experimental basis to the clinic real-time supervision of HIFU.
出处 《中华超声影像学杂志》 CSCD 2005年第3期222-224,共3页 Chinese Journal of Ultrasonography
基金 国家杰出青年基金资助项目(30325027)
关键词 高强度聚焦超声治疗 HIFU 超声声像图 凝固性坏死 靶区 生物学焦域 治疗过程 辐照 大小 变化 Ultrasonic therapy Ultrasonography High intensity focused ultrasound
  • 相关文献

参考文献7

  • 1刘宝琴,熊树华,王智彪,李发琪,伍烽,马平,涂静秋,龚小波.超声对高强度聚焦超声生物学焦域的监控研究[J].中华超声影像学杂志,2002,11(11):687-689. 被引量:26
  • 2熊树华,刘宝琴,胡凯,李发琪,田耘博,杜永洪,马平,王智彪.肝脏HIFU治疗剂量学初步研究[J].中国超声医学杂志,2002,18(8):572-574. 被引量:9
  • 3冯若 张樯 许坚毅.超声无创外科“切除”肿瘤的计量学研究[J].声学技术,2002,21:233-234.
  • 4Rowland IK, Rivens I, Chen L, et al. MRI study hepatic tumors following high intensity focused ultrasound surgery. Br J Radiol,1997,70:144-153.
  • 5Vaezy S, Shi X, Martin RW, et al. Real-time visualization of high intensity focused ultrasound treatment using ultrasound imaging.Ultrasound Med Biol,2001,27:33-42.
  • 6Sibille A, Prat F, Chapelon JY, et al. Chatacterization of extracorporeal ablation of normal and tumor-bearing liver tissue by high intensity focused ultrasound. Ultrasound Med Biol, 1993, 19.:803-813.
  • 7ter Haar GR,Sinnett D,Rivens I. High intensity focused ultrasound:a surgical technique for the treatment of discrete liver tumors. Phys Med Biol, 1989,34:1743-1750.

二级参考文献3

共引文献32

同被引文献65

引证文献7

二级引证文献32

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部