摘要 目的检测5种灭菌器对牙科手机的灭菌效果。方法用酶联免疫吸附(ELISA)法检测5种压力蒸汽灭菌器灭菌后的HBsAg人为污染的牙科手机,并对5种灭菌器灭菌后的临床污染的牙科手机进行无菌检测,同时用嗜热脂肪杆菌芽胞检测5种压力蒸汽灭菌器的灭菌效果。结果5种灭菌器生物监测均合格,5种压力蒸汽灭菌器杀菌效果都可达100%,其中2种灭菌器灭菌的所有牙科手机HBsAg抗原性检测均为阴性,其他3种灭菌器灭菌的牙科手机各有1支HBsAg抗原性检测为阳性。结论压力蒸汽灭菌器效果可靠,乙肝病毒是否灭活,还需做其他检测。牙科手机应严格遵守灭菌保养流程,灭菌后应洁净存放。 Objective To evaluate the effect of sterilization of five kinds of autoclaves on dental handpieces. Methods ELISA was used to evaluate the effect of sterilization of five kinds of autoclaves on dental handpieces which were contaminated by HBsAg in lab. Sterility testing was used to evaluate the effect of sterilization of five kinds of autoclaves on dental handpieces which were contaminated in clinic. At the same time, Geobacillus sterarothermophilus was used to assess the effect of sterilization of five kinds of autoclaves.Results The results of bio-indicators in the five kinds of autoclaves were all negative, and the killing rates were all 100%. All dental handpieces sterilized by two of five kinds of autoclaves did not exhibit HBsAg antigenicity, but there was one of 30 samples exhibited HBsAg antigenicity after sterilized by each of the three autoclaves.Conclusion The effect of sterilization of the tested autoclaves is reliable. Other tests need to be done in order to confirm whether HBV can be destroyed by autoclave.
出处 《北京口腔医学》 CAS 2005年第2期99-101,共3页 Beijing Journal of Stomatology
关键词 牙科手机 灭菌 效果 灭菌器 酶联免疫吸附 Dental handpieces Sterilization Effect
  • 相关文献

参考文献10

  • 1Matsuyama M, Usami T, Masuda K , et al. Prevention of infection in dental procedures. J Hosp Infect, 1997,35(1):17-25.
  • 2中华人民共和国卫生部.消毒技术规范[M].北京:卫生部卫生法制与监督司,2002.195.
  • 3No authors listed. Infection control recommendations for the dental office and the dental laboratory. ADA Council on Scientific Affairs and ADA Council on Dental Practice. J Am Dent Assoc, 1996, 127(5):672-680.
  • 4McCarthy GM, Koval JJ, MacDonald JK. Compliance with recommended infection control procedures among Canadian dentists: results of a national survey. Am J Infect Control, 1999, 27(5): 377-384.
  • 5McCarthy GM, MacDonald JK. Improved compliance with recommended infection control practices in the dental office between 1994 and 1995. Am J Infect Control, 1998, 26(1): 24-28.
  • 6Leonard DL, Charlton DG. Performance of high-speed dental handpieces subjected to simulated clinical use and sterilization. J Am Dent Assoc, 1999, 130(9):1301-1311.
  • 7赵昕霞,曹春菱,刘集.牙科手机经高压蒸气消毒的保养程序探讨[J].中国口腔种植学杂志,2000,5(4):187-188. 被引量:1
  • 8Andersen HK , Fiehn NE, Larsen T. Effect of steam sterilization inside the turbine chambers of dental turbines. Oral Surg Oral Med Oral Pathol Oral Radiol Endod, 1999, 87(2): 184-188.
  • 9Christensen GJ. The high-speed handpiece dilemma. J Am Dent Assoc. 1999 , 130(10): 1494-1496.
  • 10Lloyd L, Burke EJ, Cheung SW. Handpiece asepsis: a survey of the attitudes of dental practitioners. Br Dent J.1995,178(1): 23-27.

二级参考文献4

  • 1沈丽娟 张寿华 等.口内牙科手机消毒实施现状的调查分析[J].中华临床护理选萃,2000,:451-451.
  • 2徐琨 巨惠琴.口腔科院内感染控制存在的问题及改进措施[J].中华临床护理选萃,2000,:454-454.
  • 3曹健 张寿华.对病人进行口腔健康教育是21世纪护理工作者的责任[J].中华临床护理选萃,2000,:450-450.
  • 4古文珍 滕伟.口腔门诊实行“一人一机”的护理[J].中华临床护理选萃,2000,:458-458.

共引文献18

同被引文献44

引证文献5

二级引证文献25

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部