摘要
目的:用Tono-Pen眼压计测量眼球各部位眼压值。方法:正常人及无角膜病变的青光眼患者61例99眼,用Tono-Pen眼压计依次测量角膜中心、角膜缘及颞上侧角膜缘后4mm巩膜壁眼压值。结果:角膜中心眼压为(2.52±0.74)kPa,角膜缘为(2.63±0.79)kPa,巩膜壁为(4.46±1.49)kPa。角膜中心和角膜缘眼压差异无统计学意义,而巩膜壁眼压高于角膜中心。角膜中心及巩膜壁眼压呈正相关(r=0.542,P=0.000)。以Y为角膜中心眼压值,X为巩膜壁眼压值,则当X≤2.793kPa,直线回归方程为Y=1.824+0.104X;当X>2.793kPa(高眼压)时,Y=1.493+0.359X。结论:使用Tono-Pen眼压计测量眼压时,巩膜壁眼压值可为真实眼压值提供可靠参考;角膜任何部位均可测得准确眼压值。
Tono-Pen applanation tonometer; intraocular pressure; sclera; central cornea; corneal limbus Abstract Aim : To explore the difference in intraocular pressure(IOP) on different part of the eye ball measured by Tono-Pen applanation tonometer. Methods: IOP on central cornea,limbus and sclera of 61 cases(99 eyes) were measured by Tono-Pen applanation tonometer. Results: IOP on central cornea, limbus and sclera were (2. 52±0. 74 ) kPa, ( 2.63 ± 0.79) kPa, (4.46 ±1. 49 ) kPa, respectively. There was no statistical difference between the IOP on central cornea and the limbus(P 〉0.05). IOP on sclera was higher than that on corneal center,and they were positively correlated (r=0.542,P=0.000). The linear equation was Y= 1. 824 +0. 104X(X≤2.793 kPa) and Y= 1.493 +0.359(X 〉 2. 793 kPa) (Y: IOP on centralcornea;X: IOP on sclera). Conclusion :IOP on sclera measured by Tono-Pen is reliable; IOP measured by Tono-Pen at any part of cornea is accurate.
出处
《郑州大学学报(医学版)》
CAS
北大核心
2006年第1期150-151,共2页
Journal of Zhengzhou University(Medical Sciences)
关键词
Tono—Pen眼压计
眼压
巩膜壁
角膜中心
角膜缘
Tono-Pen applanation tonometer
intraocular pressure
sclera
central cornea
corneal limbus