眼球不同部位Tono-Pen眼压计测量结果比较 被引量:2

Comparison of intraocular pressures at different parts of eyeballs by Tono-Pen applanation tonometer
在线阅读 下载PDF
导出
摘要 目的:用Tono-Pen眼压计测量眼球各部位眼压值。方法:正常人及无角膜病变的青光眼患者61例99眼,用Tono-Pen眼压计依次测量角膜中心、角膜缘及颞上侧角膜缘后4mm巩膜壁眼压值。结果:角膜中心眼压为(2.52±0.74)kPa,角膜缘为(2.63±0.79)kPa,巩膜壁为(4.46±1.49)kPa。角膜中心和角膜缘眼压差异无统计学意义,而巩膜壁眼压高于角膜中心。角膜中心及巩膜壁眼压呈正相关(r=0.542,P=0.000)。以Y为角膜中心眼压值,X为巩膜壁眼压值,则当X≤2.793kPa,直线回归方程为Y=1.824+0.104X;当X>2.793kPa(高眼压)时,Y=1.493+0.359X。结论:使用Tono-Pen眼压计测量眼压时,巩膜壁眼压值可为真实眼压值提供可靠参考;角膜任何部位均可测得准确眼压值。 Tono-Pen applanation tonometer; intraocular pressure; sclera; central cornea; corneal limbus Abstract Aim : To explore the difference in intraocular pressure(IOP) on different part of the eye ball measured by Tono-Pen applanation tonometer. Methods: IOP on central cornea,limbus and sclera of 61 cases(99 eyes) were measured by Tono-Pen applanation tonometer. Results: IOP on central cornea, limbus and sclera were (2. 52±0. 74 ) kPa, ( 2.63 ± 0.79) kPa, (4.46 ±1. 49 ) kPa, respectively. There was no statistical difference between the IOP on central cornea and the limbus(P 〉0.05). IOP on sclera was higher than that on corneal center,and they were positively correlated (r=0.542,P=0.000). The linear equation was Y= 1. 824 +0. 104X(X≤2.793 kPa) and Y= 1.493 +0.359(X 〉 2. 793 kPa) (Y: IOP on centralcornea;X: IOP on sclera). Conclusion :IOP on sclera measured by Tono-Pen is reliable; IOP measured by Tono-Pen at any part of cornea is accurate.
出处 《郑州大学学报(医学版)》 CAS 北大核心 2006年第1期150-151,共2页 Journal of Zhengzhou University(Medical Sciences)
关键词 Tono—Pen眼压计 眼压 巩膜壁 角膜中心 角膜缘 Tono-Pen applanation tonometer intraocular pressure sclera central cornea corneal limbus
  • 相关文献

参考文献9

  • 1Frenkel RR,Hong YJ,Shin DH.Comparison of the Tono-Pen to the Goldmann applanation tonometer.Arch Opthalmol,1988,106(6):750.
  • 2Kao SF,Lichter PR,Bergstrom TJ,et al.Clinical comparison of the Oculab Tono-Pen to the Goldmann applanation tonometer.Ophthalmology,1987,94(12):1541.
  • 3Hines MW,Jost BF,Fogelman KL.Ocular Tono-pen,Goldmann applanation tonometry and pneumatic tonometry for intraocular pressure assessment in gas-filled eyes.Am J Ophthalmol,1988,106 (2):174.
  • 4Boothe WA,Lee DA,Panek WC,et al.The Tono-Pen.A manometric and clinical study.Arch Opthalmol,1988,106 (9):1214.
  • 5Tonnu PA,Ho T,Newson T,et al.The influence of central corneal thickness and age on intraocular pressure measured by pneumotonometry,non-contact tonometry,the Tono-Pen XL,and Goldmann applanation tonometry.Br J Ophthalmol,2005,89 (7):851.
  • 6Garzozi H J,Chung HS,Lang Y,et al.Intraocular pressure and photorefractive keratectomy:a comparison of three different tonometers.Cornea,2001,20(1):33.
  • 7Schipper I,Senn P,Oyo-Szerenyi K,et al.Central and peripheral pressure measurements with the Goldmann tonometer and Tono-Pen after photorefractive keratectomy for myopia.J Cataract Refract Surg,2000,26 (6):929.
  • 8Park H J,Uhm KB,Hong C.Reduction in intraocular pressure after laser in situ keratomileusis.J Cataract Refract Surg,2001,27 (2):303.
  • 9Sullivan-Mee M,Pham F.Correspondence of Tono-Pen intraocular pressure measurements performed at the central cornea and mid-peripheral cornea.Optometry,2004,75 (1):26.

同被引文献10

引证文献2

二级引证文献6

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部