期刊文献+

覆膜电极溶解氧测定仪示值误差的探讨 被引量:12

Discussing about indication error of film electrode dissolved oxygen meter
在线阅读 下载PDF
导出
摘要 提出了一种在空气中测量覆膜电极溶解氧测定仪示值误差的方法。首先分析了相关的理论基础,并且介绍了覆膜电极溶解氧测定仪及其示值误差的检测原理,分析了JJG 291-1999《覆膜电极溶解氧测定仪》中示值误差检定方法的不足之处,提出了采用操作简单,准确可靠的空气测量法,能够实现覆膜电极溶解氧测定仪示值误差的检定。 In this article the authors proposed a method for measures indication error of film electrode dissolved oxygen meter:air calibration. Firstly they analyzed the basle theory and presented a description of this method. They also analyzed the method of the JJG291-1999 (Film Electrode Dissolved Oxygen Meter).Tne results show that this method (air calibration) is practical for measuring indication error of film electrode dissolved oxygen meter.
作者 陈炜庆
出处 《中国测试技术》 2006年第5期72-73,76,共3页 CHINA MEASUREMENT & TESTING TECHNOLOGY
关键词 覆膜电极 溶解氧 溶解氧测定仪 示值误差 检定 Film electrode Dissolved oxygen Dissolved oxygen meter Indication error Verification
  • 相关文献

参考文献6

  • 1JJG 291-1999,覆膜电极溶解氧测定仪[S].
  • 2JJG 291-1982,复膜电极溶解氧测定仪试行检定规程[S].
  • 3SC/T 7006-2001,溶解氧测定仪[S].
  • 4GB 11913-1989,水质溶解氧的测定电化学探头法[S].
  • 5CB 7489-1987,水质 溶解氧的测定 碘量法[S].
  • 6ISO 5814:1990,Water Quality-Determination of Dissoled Oxygen-Electrochemical Probe Method[S].