期刊文献+

PRVC和VC(A/C)治疗AECOPD呼吸衰竭时的人机同步性研究 被引量:5

Research on mechanical synchronism of PRVC and VC(A/C) for respiratory failure AECOPD patients
在线阅读 下载PDF
导出
摘要 目的研究慢性阻塞性肺疾病急性加重期(AECOPD)患者呼吸衰竭时初始采用压力调节容量控制通气(PRVC)治疗的人机同步性。方法将10例并发呼吸衰竭需行机械通气治疗的AECOPD患者随机分为试验组和对照组。试验组以PRVC进行治疗,对照组以容量控制的辅助/控制通气(VC(A/C))治疗,分别于机械通气治疗1、2、3、4h观察两组并记录两组于每小时所发生的人机对抗次数。结果在同一观察时间,试验组所发生的人机对抗次数多于对照组(P<0.05),而4h内所发生的人机对抗总次数也明显多于对照组(P<0.05)。结论在AECOPD患者呼吸衰竭行机械通气治疗时初始采用PRVC模式时的人机同步性不及VC(A/C)模式,故此时采用PRVC模式治疗应慎重。 Objective To study the mechanical synchronism of PRVC (Pressure Regulation Volume Control) on respiratory failure of AECOPD (acute exacerbation of chronic obstructive pulmonary disease) patients. Methods Ten COPD patients with respiratory failure who should receive mechanical ventilation treatment were divided into two groups randomly: experimental group which received PRVC and the control group which experienced VC ( A/C ) ( assistant/control of volume control). After beginning of the mechanical ventilation, times of asynchronous ventilation of each hour were recorded during the first 4-hour ventilation. Results The times of asynchronous ventilation from the experimental group were more than that in the controls: P 〈 0. 05. Conclusions The mechanical synchronism of PRVC for respiratory failure AECOPD patients was not better than that of VC(A/C) at the beginning of ventilation treatment, it should be careful to treat this kind of respiratory failure by applying PRVC as the first mode of mechanical ventilation.
出处 《临床肺科杂志》 2007年第4期338-339,共2页 Journal of Clinical Pulmonary Medicine
  • 相关文献

参考文献5

二级参考文献22

  • 1朱蕾.有创-无创“序贯”机械通气治疗慢性阻塞性肺疾病呼吸衰竭[J].中国呼吸与危重监护杂志,2003,2(4):198-199. 被引量:51
  • 2朱蕾,钮善福,蔡映云,徐风平,李燕芹,李善群.经面罩机械通气治疗慢性阻塞性肺病呼吸衰竭患者[J].上海医科大学学报,1996,23(5):347-350. 被引量:23
  • 3Holt SJ , Sanders RC. An evaluation of automode a computer - controlled ventilator mode, with the Siemens Servo 300A ventilator using a porcine model [J]. Respir Care, 2001,46(1). 26-35.
  • 4Murray .IF, Matthay MA, Luce JM, et al. An expanded definition of the adult respiratory distress syndrome[J]. Am Rev Respir Dis, 1998,138:720 - 723.
  • 5Kopp R,Kuhlen R,Max M. Evidence-based medicine in the therapy of the acute respiratory distress syndrome [ J ]. Intens Care Med, 2002, 28(3) :244 - 256.
  • 6Riverso P, Bemardi PL, Corsa D, et al. A comparison of ventilation techniques in ARDS. Volume controlled vs pressure regulated volume control [ J]. Minerva Anesthesiol, 1998,64(78) : 339 - 342.
  • 7Pierce JD,Gilliland E,Smith - Blair N, et al. Effects of volume control,pressure control, and pressure - regulated volume control on cardiopulmonary parameters in a normal rat lung [J]. Mil Med, 1998,163(9) :625 - 630.
  • 8Kocis KC, Dekeon MK, Rosen HK, et d. Pressure - regulated volume control vs volume control ventilation in infants after surgery for congenital heart disease [J]. Pediatr Cardiol, 2001,22(3) :233 - 236.
  • 9中华医学会呼吸病学分会.慢性阻塞性肺疾病(COPD)诊治规范[J].中华结核和呼吸杂志,2002,:25-460,443.
  • 10王辰.慢性阻塞性肺疾病急性发作期机械通气治疗[A].刘又宁主编.机械通气与临床[C].北京:科学技术出版社,1990.2347-349.