摘要 目的:探讨C型臂X线透视判断上中胸椎椎弓根螺钉置入位置的准确性。方法:选用10具成年防腐保存的脊柱标本(T1~T6),根据椎弓根水平面倾斜角用手锥建立螺钉位置偏内(38个)、偏外(36个)、居中(46个)3种模型,分别置入导针和椎弓根螺钉后依次透视正位、侧位、椎弓根轴位、导针轴位、椎弓根螺钉轴位,根据导针和椎弓根钉与椎弓根投影的相互位置关系判断导针和椎弓根钉位置,同时行CT扫描观察椎弓根螺钉位置以及椎弓根内外侧皮质穿透情况,根据CT扫描结果判断C型臂X线图像对椎弓根螺钉置入监测的准确率。结果:120个椎弓根正位透视10枚偏内,其中5枚实际位置良好,未发现偏外者,正确率为35%;侧位透视不能判断偏内及偏外错误;椎弓根轴位透视将5枚居中判为偏内,准确率为95.9%;导针和椎弓根钉轴位透视均将4枚偏外模型判为居中,准确率为96.7%。结论:C型臂X线透视时椎弓根轴位和导针轴位图像能较准确判断椎弓根钉导针位置。 Objective:To explore the reliability of C-arm fluoroscopy in judging the accuracy of pedicle screws' position applied in the upper-middle thoracic region.Method:10 adult cadaveric thoracic samples(T1- T6) were selected.According to the incline angle of thoracic pedicle in the axial plane,3 kind models were established. After the surgical probe and pedicle screw were placed respectively,the fluoroscopic imaging of A-P view,lateral view,axis view of pedicle,axis view of probe and axis view of pedicle screw were observed individually.According to the projection of the probe and the pedicle screw,the relative position of the probe and the pedicle screw were determined.After that,CT scan was performed to confirm the position of screws and evaluate the penetration of the inside and outside cortexes of the pedicles.By compared with the result of CT scaning,the accuracy of the resuh of the C- arm fluoroscopy was determined.Result:Out of the 10 screws which judged inwardly under the AP view,5 screws were actually in right position,no screw was found out-wardly.The accuracy rate reached 35%.The inward or outward deviation of the screws could not be determined by the lateral view.5 screws in right position were judged to be inward deviation by the axis view of pedicle, the accuracy rate was 95.9%.4 outward deviation screws were judged to be central position by the axis view of probe and axis view of pedicle screw,the accuracy rate was 96.7%.Conclusion:By means of C-arm fluoroscopy,the axis view of probe and the axis view of pedicle screw can accurately reflect the position of the pedicle screw.
出处 《中国脊柱脊髓杂志》 CAS CSCD 2008年第2期142-145,共4页 Chinese Journal of Spine and Spinal Cord
关键词 胸椎 C型臂X线机 椎弓根 螺钉 Upper-middle thoracic pedicle C-shaped arm X-ray fluoroscopy Pedicle Screw
  • 相关文献

参考文献10

  • 1Kim YJ,Lenke LG, Bridwell KH, et al. Free hand pedicle screw placement in the thoracic spine:is it saft [J]? Spine, 2004,29 (3) : 333-342.
  • 2Suk SS,Kim WJ,Lee SM,et al.Thoracic pedicle screw fixation in spinal deformities,are they really safe[J].Spine,2001,26 (18) : 2049-2057.
  • 3吕厚山.脊柱内固定学[M].第二版.北京:中国医药科技出版社,2001.261-263.
  • 4陈迎春,郭昭庆,徐万鹏.腰椎椎弓根导针轴位透视监测的实验研究[J].中华医学杂志,2006,86(19):1309-1312. 被引量:7
  • 5Sandeep P, Datir MS, Sajal R. Morphomttnc study of the thoracic vertebral pediclt in an India population [J].Spine, 2004,29(11):1174-1181.
  • 6Lehman RA Jr,Kuklo TR.Use of the anatomic trajectory for thoracic pedicle screw salvage after failure/violation using the straight-forward technique: a biomechanical analysis[J].Spine, 2003,28 ( 18 ) : 2072-2077.
  • 7毛克亚,王岩,张永刚,刘郑生,肖嵩华,刘保卫.胸椎椎弓根螺钉置入位置的CT评价[J].中国脊柱脊髓杂志,2005,15(4):222-224. 被引量:21
  • 8Gu Y,Ebraheim NA,Yeasting RA.Identification of the anterior and posterior portions of the lateral pedide cortek by roentgenograms in pedicle screw fixation[J].Surg Radio Anant, 2002,24(3) :313-318.
  • 9Weinstein JN, Spratt KF, Spengler D, et al. Spinal pedicle fixation .. reliabilility and validity of roentgenogram -based assessment and surgical factors on successful screw placement [J].Spine, 1998,13(10) : 1012-1018.
  • 10Carbone JJ,Tortolani PJ,Quartararo LG. Fluoroscopically assisted pedicle screw fixation for thoracic and thoracolumbar injuries technique and short-term complications [J].Spine, 2003,28( 1 ) :91-97.

二级参考文献25

  • 1Lehman RA,Kuklo TR.Use of the anatomic trajectory for thoracic pedicle screw salvage after failure/violation using the straight-forward technique:a biomechanical analysis[J].Spine,2003,28 (18) :2072-2077.
  • 2Kim YJ, Lenke LG, Cho SK,et al. Comparative analysis of pedicle screw versus hook instrumentation in posterior spinal fusion of adolescent idiopathic scoliosis [J].Spine,2004,29(3):2040-2048.
  • 3Kim YJ,Lenke LG,Bridwell KH,et al. Free hand pedicle screw placement in the thoracic spine:is it safe[J]?Spine,2004,29(3):333-342.
  • 4Xu R,Ebraheim NA,Ou Y,et al. Anatomic considerations of pedicle screw placement in the thoracic spine:Roy-Camille technique versus open-lamina technique[J].Spine,1998,23 (9):1065-1068.
  • 5Kuklo TR,Lenke LG,O′Brien MF,et al. Accuracy and efficacy of thoracic pedicle screws in curves more than 90 degree[J].Spine, 2005,30 ( 2 ):222-226.
  • 6Parent S,Labelle H,Skalli W,et al. Thoracic pedicle morphometry invertebrae from scoliotic spines [J].Spine,2004,29 (3):239-248.
  • 7Kim KD,Johnson JP,Bolch O,Masciopinto JE.Computer-assited thoracic pedicle screw placement:an in vitro feasibility study [J].Spine, 2001,26 (4) :360-364.
  • 8Suk SS,Kim WJ,Lee SM,et al.Thoracic pedicle screw fixation in spinal deformities:are they really safe[J]?Spine,2001,26(18) :2049-2057.
  • 9Sandeep P,Datir MS,Sajal R. Morphometric study of the thoracic vertebral pedicle in an Indian population[J].Spine,2004,29(11 ):1174-1181.
  • 10Liljenqvist UR,Link TM,Halm HFH.Morphometric analysis of thoracic and lumbar vertebrae in idiopathic scoliosis[J].Spine,2000,25 (10):1247-1253.

共引文献26

同被引文献42

引证文献4

二级引证文献3

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部