摘要
利用全最小二乘法,对丙酮中所含红外光谱重叠严重的苯及甲苯所测数据,自编程序,在计算机上进行处理,所得结果相对误差均小于6%,优于最小二乘法.
A new method based on the total least squares is proposed. Made use of it combined with infrared spectroscopy to determine benzene and toluene in acetone whose infrared spectrums are overlapped.The results display that the method of total least squares is better than the method of the least squares.
出处
《郑州轻工业学院学报》
1997年第2期55-58,共4页
Journal of Zhengzhou Institute of Light Industry(Natural Science)
基金
河南省自然科学基金
关键词
最小二乘法
红外测量
丙酮
苯
甲苯
least squares methods
infrared spectrometry
acetone
benzene
toluene