期刊文献+

根尖定位仪Raypex~ 5测量准确性实验研究 被引量:6

Evaluation of the measurement accuracy of apex locator Raypex~ 5.
在线阅读 下载PDF
导出
摘要 目的:研究根尖定位仪Raypex 5测量特点。方法:采用生理盐水和单根管离体牙构建实验研究模型,使用Raypex 5在离体牙根管内存在0.9%生理盐水时进行根管长度测量,记录测量值和实际值之间的差值(DL)和其分布。对差值数值采用单因素方差分析进行统计学分析。结果:使用第1~3绿色数码条测量准确率在根管真实长度±0.50 mm内分别为55%,65%和70%,DL均值分别为-0.42±0.48 mm,-0.35±0.44 mm和-021±0.41 mm。单因素方差分析显示3组DL值间无显著性差异(P>0.05)。结论:结合测量准确率和DL值均值,在本实验条件下Raypex 5第三个绿色显示条对应测量结果最接近根管真实长度。 Objective: To investigate the operation property of apex locator Raypex 5. Methods : We constructed evaluation model with agar and single - canal extracted tooth, and recorded the distance between measured and actual length as DL value by using Raypex 5 with 0.9% NaC1 inside the canal. The One - Way ANOVA test was applied to analyze the result. Results: The accuracy of measurement was 55% for the first green bar, 65% for the second and 70% for the third. The mean value of DL were - 0.42 ± 0.48mm, - 0.35 ± 0.44ram and - 021 ± 0.41mm respectively. No significant difference was found between them( P 〉 0.05 ). Conclusion: The result obtained by the third green bar was closest to the true length regarding the accuracy and mean value of t in this laboratory study.
出处 《口腔医学研究》 CAS CSCD 2008年第6期695-696,共2页 Journal of Oral Science Research
关键词 根尖定位仪 根管长度 测量准确性 Apex locator Root canal length Accuracy of measurement
  • 相关文献

参考文献6

  • 1Gordon MPJ, Chandler NP. Electronic apex locators[J].Int Endod J, 2004, 37 (7) : 425 - 437
  • 2陈惠萍,王茜,王双.Root ZX根尖定位仪测量恒牙工作长度的体内准确性研究[J].口腔医学研究,2006,22(6):695-695. 被引量:4
  • 3Kim E, Lee SJ. Electronic apex locator [ J ]. Dent Clin North Am, 2004, 48(1):35-54
  • 4Goldberg F, Marroquin BB, Frajlich S, Dreyer C. In vitro evaluation of the ability of three apex locators to determine the working length during retreatment[J].J Endod, 2005, 31 (9):676 - 678
  • 5Kaufman AY, Keila S, Yoshpe M. Accuracy of a new apex lotator: an in vitro study [J]. Int Endod J, 2002, 35(2): 186 -192
  • 6Wrbas KT, Ziegler AA, Altenburger M J, et al. In vivo comparison of working length determination with two electronic apex locatots [J]. Int Endod J, 2007, 40(2): 133 -138

共引文献3

同被引文献29

  • 1顾永春,皮昕,周培刚,刘淑香.3803个恒牙侧副管的解剖形态研究[J].临床口腔医学杂志,2004,20(6):334-338. 被引量:16
  • 2朱习华,董红,龙明生.ROOT ZX根管测量仪测量根管长度的临床分析[J].北京口腔医学,2006,14(1):63-63. 被引量:7
  • 3潘广嗣,吉建新,高平.拔牙实测法评价Neosono根尖定位仪的准确性[J].广州医学院学报,2006,34(4):29-31. 被引量:3
  • 4Goldberg F, Marroquin BB, Frajlich S. In vitro evaluation of the ability of three apex locators to determine the workong length during retreatment [J]. J Endod, 2005, 31(9): 676-678.
  • 5Kaufman AY, Keial S, Yoshpe M. Accuracy of a new apex locator: an in vitro study [J].Int EndodJ, 2002, 35(2): 186-192.
  • 6Wrbas KT,Ziegler AA, Altenburger MJ, et al. In vivo comparison of working length determination with two electronic apex locators [J]. Int Endod J, 2007, 40(2) : 133 - 138.
  • 7Moshonov J,Slutzky-Goldberg I.Apex locators:update and prospects for the future.Int J Comput Dent,2004,7(4):359-370.
  • 8Donnelly JC.A simplified model to demonstrate the operation of electronic root canal measuring device.J Endod,1993,19(11):579-580.
  • 9Herrera M,Abalos C,Planas AJ,et al.Influence of apical constriction diameter on Root ZX apex locator precision.J Endod,2007,33(8):995-998.
  • 10Czerw RJ,Fulkerson MS,Donnelly JC.An in vitro test of a simplified model to demonstrate the operation of electronic root canal measuring devices.J Endod,1994,20(12):605-606.