期刊文献+

HPGeγ谱仪效率曲线的拟合 被引量:6

Fitting of Efficiency Curves for HPGe γ Spectrometer
在线阅读 下载PDF
导出
摘要 对HPGeγ谱仪探测效率在39.52-1 408 keV能区进行分段多项式拟合及整个区域一次拟合,并对两种拟合方式进行了比较。结果表明:在实验能区内,整体一次多项式拟合的计算相对简单,但分段拟合获得数据更为准确。用两种拟合函数对标准源及样品的活度进行分析,分段拟合方式的结果优于整体一次多项式拟合。 The energy peak efficiency of HPGe gamma-ray spectrometer was obtained by he piecewise polynomial fitting and the full-energy polynomial fitting in the 39.52~1 408 keV energy range,and the two types of curves fitting were compared with each other.It was shown that although calculating process of the full-energy polynomial fitting was easier,dates acquired from the piecewise polynomial fitting was more accurate.Besides,the activity of standard sources and samples were analyzed by two kinds of curves fitting,the fitting results of the later was superior to the former.
出处 《辐射防护通讯》 2009年第6期35-37,共3页 Radiation Protection Bulletin
关键词 HPGEΓ谱仪 效率刻度 拟合 HPGe γ spectrometer Efficiency Piecewise polynomial fitting
  • 相关文献

参考文献3

二级参考文献11

  • 1International Atomic Energy Agency. X-ray and Gamma-ray Standards for Detector Calibration:IAEA-TECDOC-619[R]. Vienna:IAEA,1991.
  • 2Chmid GJS, Chastelet RM, Laymon CM, et al.Correction Factors for γ-ray Relative Intensities in the 66 Ga Radioisotope[ J ]. Nucl Phys, 1996,A (607) :139-151.
  • 3Kennett TJ, Prestwich WV, Tsai JS. The ^14N(n,γ)^15 N Reaction as Both an Intensity and Energy Standard[J].Nucl Instrum Methods,1986, A(249) : 366 - 378.
  • 4Spits AMJ, Kopecky J. The Reaction 3s Cl(n, γ)^36C1 Studied With Non-polarized and Polarized Thermal Neutrons [J]. Nucl Phys, 1976,A (264) :63-92.
  • 5Ruy JFAG, Endt PM. Investigation of the ^48Ti(n,γ)^49Ti Reaction[J]. Nucl Rhys, 1983,A (407) :60-76.
  • 6Raman S, Yonezawa C, Matsue H, et al. Efficiency Calibration of a Ge Detector in the 0. 111. 0 MeV[J]. Nucl Instrum Methods, 2000,A(454) :389-402.
  • 7Singh BP, Evans HC. Relative Efficiency of Ge(Li) Gammy-ray Detectors From 0. 5 to 12 MeV[J]. Nucl Instrum Methods, 1971, ( 97 ) :475-482.
  • 8Zifderhand F, Jansen P, Alderliesten C, et al.Detector-efficiency Calibration for High-energy Gamma-rays[J]. Nucl Instrum Methods, 1990,A (286) :490-496.
  • 9Tabata T, Okano K. Gamma-rays From the 7.56 MeV Level in O^15 [J]. J Phys Soc Jpn,1960,15(9) :1552-1555.
  • 10Kis Z, Fazekas B, Ostor J, et al. Comparison of Efficiency Functions for Ge Gamma-ray Detectors in a Wide Energy Range[J]. Nucl Instrum Methods, 1998 ,A(418) : 374- 386.