摘要
目的 探讨80 kV管电压256层螺旋CT主动脉造影降低射线剂量的可行性.方法 对62例疑似主动脉疾病的患者进行主动脉CT造影成像,分为常规管电压扫描组(120 kV扫描组)和低管电压扫描组(80 kV扫描组),每组31例.然后对两组图像信号噪声比、对比度噪声比和有效剂量进行评估.结果 120与80 kV2个扫描组的图像信号噪声比分别为(35.92±5.04)和(33.95±8.30),对比度噪声比分别为(30.32±4.78)和(28.71±7.96),差异均无统计学意义(t=1.131、0.964,P>0.05).而二组的有效剂量分别为(14.28±0.96)和(9.72±0.81)mSv,差异有统计学意义(t=20.12,P<0.001).结论 同120 kV管电压比较,80 kV管电压主动脉CT造影在没有降低图像质量的情况下,射线剂量降低约31.9%,同时降低造影剂剂量50%.
Objective To evaluate the feasibility of 80 kV tube kilovoltage in aortic artery CT angiography with 256-slice CT. Methods A total of 62 patients undergoing aortic artery CTA were enrolled into this study and divided into conventional tube kilovoltage (120 kV, n = 31 ) and low tube kilovoltage (80 kV, n = 31 ). The signal-to-noise ratio (SNR) , contrast-to-noise ratio (CNR), volume CT dose index (CTDIvol) and effective dose (E) were evaluated, respectively. Results The mean image SNR was ( 35.92 ± 5.04) and ( 33.95 ± 8. 30) for conventional tube kilovoltage and low tube kilovoltage,respectively, with no significant difference (t = 1. 131, P =0. 263). The mean image CNR was (30. 32 ±4.78) and (28.71 ± 7.96 ) for conventional tube kilovoltage and low tube kilovoltage, respectively, with no significant difference ( t = 0. 964, P = 0. 339 ). The average effective dose ( E ) was ( 14. 28 ± 0.96 )mSv and (9. 72± 0. 81 )mSv for conventional tube kilovoltage and low tube kilovoltage, respectively, with significant difference ( t = 20. 12, P < 0. 001 ). Conclusions 80 kV tube kilovoltage aortic artery CTA can reduce radiation dose by 31.9% , and contrast dose 50% ,and maintain image quality compared with 120 kV tube kilovoltage.
出处
《中华放射医学与防护杂志》
CAS
CSCD
北大核心
2010年第6期740-743,共4页
Chinese Journal of Radiological Medicine and Protection