摘要 针对设计与研制的一台活体小动物X射线显微CT成像系统(Micro-CT),利用标准的细丝模具成像实验,测量该系统的调制传递函数(MTF)曲线,衡量整个系统的空间分辨率。该系统在其常规的成像条件下(X射线管电压为40kVp,曝光时间1s,几何放大倍数约2倍,探测器设置在2×2工作模式),测得系统的成像分辨率约11lp/mm(MTF为0.1)。测量结果表明,该系统的成像分辨率完全达到了100μm的设计要求,能提供高质量的小动物成像。 The aim of the study was to determine the spatial resolution of the small animal micro-CT system.A standard wire phantom was used to measure the modular transfer function(MTF)of the micro-CT system.On the imaging conditions with X-ray tube voltage of 40 kVp,exposure time of 1 second,detector mode of 2×2, and system magnification factor of 2,the measured system spatial resolution is ~11lp/mm at 10%MTF,which is about 91 microns.The results showed that the small animal micro-CT that we developed has met our design goal to achieve the spatial resolution under 100 microns.
出处 《核电子学与探测技术》 CAS CSCD 北大核心 2011年第3期293-296,318,共5页 Nuclear Electronics & Detection Technology
基金 中国科学院百人计划择优支持项目(26050601)
关键词 显微CT 调制传递函数(MTF)曲线 空间分辨率 micro-CT modular transfer function(MTF) spatial resolution
  • 相关文献

参考文献5

二级参考文献5

  • 1陈成杰.光电倍增管[M].北京:原子能出版社,1998..
  • 2[1]Keepin G R.Nuclear Kinetics,First Edition[M].USA,Addison-Wesley Publishing Company,1965.166.
  • 3[2]Orndoff J D.Prompt Neutron Periods of Metal Critical Assemblies[J].Nucl Sci Eng,1957,2:450~460.
  • 4[1]Wimett T F,White R H.Godiva Ⅱ-An Unmoderated Pulse-Irradiation Reactor[J].Nucl.Sci.and Eng.,1960,8:691 ~ 708.
  • 5范晓强,李兵,邓门才.用Rossi-α法测量快临界装置瞬发中子衰减常数α[J].核电子学与探测技术,2001,21(4):268-270. 被引量:2

共引文献29

同被引文献6

  • 1李兆斌 孔祥闯 余建明.数字X线摄影的调制传递函数测量方法.中国现代影像学杂志,2005,2(5):424-425.
  • 2Fujita H. ,Tsai D -Y. ,Itoh T. ,et al. A simple meth- od for determining the modulation transfer function in digital radiography[J]. IEEE Transactions on Medical Imaging. 1992, 11 ( 1 ) :34 - 39.
  • 3Yamazaki T, Nokita M, Hayashida S, et al. A meth- od to measure the pre - sampling MTF using a novel edge test device and algorithm[ C] //Medical Imaging 2004. International Society for Optics and Photonics. 2004 : 696 - 704.
  • 4K. Y., AL. K., JM. B. Computer modeling of the spatial resolution properties of a dedicated breast CT system [ J ]. Medical Physics. 2007, 34 ( 6 ) : 2059 - 2069.
  • 5Mettivier G, Russo P. Measurement of the MTF of a Cone- Beam Breast Computed Tomography laboratory scanner[J]. Nuclear Science Symposium Conference Record (NSS/MIC), 2011 IEEE. 2011, 58 (3) : 703 -713.
  • 6康立丽,邓晓刚,陈阳,杨绍洲.影像系统MTF算法实现及分析[J].北京生物医学工程,2002,21(3):179-179. 被引量:13