双探头SPECT质控回顾性分析 被引量:4

Retrospective analysis for quality control of a dual-detector SPECT
在线阅读 下载PDF
导出
摘要 根据NEMA标准对一台SPECT进行固有均匀性、旋转中心校正、四象限铅栅模型、Jaszczak系统模型、Hoffman脑3D模型测试。2003–2010年质控数据分析结果显示,有效视野(UFOV)固有积分均匀性与旋转中心均无明显差异(p>0.05),中心视野(CFOV)固有积分均匀性从2007年开始有明显差异(p<0.05)。表明仪器随着年限的延长,性能参数逐渐变差,通过质控能将参数维持在标准范围,保证图像质量以及诊断符合率。 In this work, quality control data of a SPECT in 2003-2010 were analyzed by t test. According to the NEMA standard for uniformity, the center of rotation, bar phantom, Jaszczak phantom and the Hoffman phantom were measured in the QC. Compared with the 2003 data, no significant difference was found in the UFOV intrinsic integral uniformity and center of rotation, but the CFOV intrinsic integral uniformity has been worsening since 2007 (p〈0.05), indicating the performance decrease with equipment age. By the QC measurements, we were able to ensure image quality and diagnostic accuracy of the SPECT.
出处 《核技术》 CAS CSCD 北大核心 2011年第9期701-704,共4页 Nuclear Techniques
关键词 回顾性分析 质量控制 双探头 Restrospective analysis, Quality control, Dual-detector
  • 相关文献

参考文献6

二级参考文献8

共引文献35

同被引文献33

  • 1周铭娟.对SPECT均匀性的测试与校正[J].医疗设备信息,1995,10(3):6-7. 被引量:2
  • 2钱英.SPECT探头系统的校正与调试[J].医疗设备信息,2006,21(5):40-41. 被引量:1
  • 3贺小红,贺小红.NEMA标准的参考资料、概念定义和测试设备[J].国外医学(放射医学核医学分册),1997,21(2):49-51. 被引量:5
  • 4Christian La Foug're,Walter Hundt, Nieole Briiekel,et (d. Value of PET/CT versus PET and CT performed as separate investigations in patients with Hodgkin's lymphoma[J]. Eur J Nuel Med Mol hnaging, 2006,33(12):1 417-1 425.
  • 5Xingehen Wu, Prasn I)astidar, Harem Petlnvaara,et d. Early Treat- ment Response Evaluation in Patients with Diffuse Large B-Cell Lymphoma-A Pilt Study Comparing Volumetric MRI amt PET/CT [J]. Mol Imaging Bio1,2011,13(4) :785-792.
  • 6Ronald Boellaard,Mike J O'Doherty,Woifgang A,et d. FI)G PET and PET/CT: EANM procedure guidelines fir tumour PET imaging, version 1.01J]. Eur J N ucl Med Mol Imaging, 2010,37 ( 1 ) : 181-200.
  • 7Lyrao Georgosopoulou. Wendy Tindale. Rigobe,l Klett ,et d. Routine quality control recmmendations for nuclear medicine instnmlentation [J]. Eur J Nucl Med Mol Imaging. 201 O, 37 (3) : 662-671.
  • 8Mattsson S, Hoeschen C. Radiation protection in nuclear medicine[M]. Heidelberg: Springer, 2012: 71-93.
  • 9EANM Physics Committee, Busemann Sokole E, Piachclnska A, et al. Routine quality control recommendations for nuclear medicine instrumentation[J]. Eur J Nucl Med Mol Imaging, 2010, 37(3): 662-671.
  • 10Patton JA, Turkington TG. SPECT/CT physical principles and attenuation correction[J], l Nucl Med Technol, 2008, 36(1): 1-10.

引证文献4

二级引证文献6

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部