非晶硅电子射野影像装置的性能研究 被引量:3

Study on dose-response characteristics with amorphous silicon electronic portal imaging device
在线阅读 下载PDF
导出
摘要 使用DVS(Dosimetric verifying system)软件实现对非晶硅平板探测器的影像采集,通过分析不同辐照条件下得到的射野影像的像素值,研究非晶硅电子射野影像装置(Amorphous silicon electronic portal Imagingdevice,a-Si EPID)的辐射稳定性、剂量响应线性、照射野大小的散射影响以及不同帧时间、不同电容大小等参数设置对EPID读出信号值的影响。结果显示EPID在短期内显示出极好的稳定性,伪影增加了EPID的响应信号,减少两次照射的间隔时间或者增加两次照射剂量比例可使伪影变得更加明显。a-Si EPID具有良好的剂量线性关系,随着射野的增加散射因子也随着增加。因此只要经过适当的散射修正和剂量刻度,EPID可以作为剂量仪用来显示二维剂量分布。 The object of this study is to investigate the dose-response characteristics of an amorphous silicon electronic portal imaging device (a-Si EPID). An in-housing developed DVS(Dosimetric verifying system) software was used to analysis the portal image captured in different conditions, and the relationship of pixel value with dose, field size, dose rate, readout signal variation in different binning mode, gain mode and integration time setting was investigated. The EPID showed an excellent temporal stability on short time .Ghosting effects could increase the sensitivity of the EPID and became more pronounced with decrement of time intervals between two exposures as well as with increment of dose though the EPID response is linear with dose. However, it will be ideal for dosimetry application if some corrections for scatter could be made.
出处 《辐射研究与辐射工艺学报》 CAS CSCD 2012年第3期178-182,共5页 Journal of Radiation Research and Radiation Processing
关键词 非晶硅电子射野影像装置 非晶硅平板探测器 剂量学特性 Electronic portal imaging device, A-Si EPID, Dose-response characteristics
  • 相关文献

参考文献8

  • 1Boellaard R, van Herk M, Uiterwaal H, et al. First clinical tests using a liquid-filled electronic portal imaging device and a convolution model for the verification of the midplane dose. Radiother. Oncol, 1998, 47:303-312.
  • 2Steciw S, Warkentin B, Rathee S, et al. Three-dimensional IMRT verification with flat-panel EPID. Med Phys, 2005, 32:600-612.
  • 3Awusi Kavuma, Martin Glegg, Garry Currie, et al. Assessment of dosimetrical performance in 11 Varian a-Si500 electronic portal imaging devices. Phys Med Biol, 2008, 53:6893-6909.
  • 4Winkler P, Georg D. An intercomparison of 11 amorphous silicon EPIDs of the same type: implications for portal dosimetry. Phys Med Biol, 2006, 51:4189-4200.
  • 5Homayoon Parsaei, Ellen E1-Khatib. The use of an electronic portal imaging system to measure portal dose and portal dose profiles [J]. Med Phys, 1998, 25: 1903-1909.
  • 6Louwe R J W, McDermott L N, Snoke J J, et al. The long-term stability of amorphous silicon flat panel imaging devices for dosimetry purposes[J]. Med Phys, 2004, 31:2989-2995.
  • 7Winkler P, Hefner A, Georg D. Dose-response characteristic of an a-Si EPID [J]. Med Phys, 2005, 32: 3095- 3105.
  • 8McDermott L N, Louwe R J, Sonke J J, et al. Comparison of ghosting effects for three commercial a-Si EPIDs. Med Phys, 2006, 33:2448-2451.

同被引文献12

引证文献3

二级引证文献17

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部