小儿罗叶泵的研制 被引量:1

Development of Pediatric Luo-Ye Pump
在线阅读 下载PDF
导出
摘要 目的探讨20 mL小儿罗叶泵的制作方法,并对其行耐疲劳实验,观察罗叶泵的搏动性能。方法罗叶泵由泵体、隔膜、隔膜固定圈、泵接口、出入口和人工瓣膜等组成。模具制作,采用注塑成形工艺制作辅助泵泵体和隔膜固定圈;采用冷成形热固定的方法制作罗叶泵隔膜;泵的出、入接口采用Tc-4医用钛金属制作;瓣膜采用17#CL-I(标准型)国产人工机械心脏瓣膜。对组装好的罗叶泵在模拟循环台上行耐疲劳实验。结果采用模具注塑成形的方法制作20 mL小儿罗叶泵,泵内表面光滑,泵体透明。在模拟循环台上进行测试,泵体、隔膜及接口无变形和渗漏。结论采用模具注塑成形技术可成功制作小儿罗叶泵,质量稳定,可实现批量生产,提高了制作效率。制成的罗叶泵具有良好的耐久性。 Objective To investigate the manufacture method of 20 mL pediatric Luo-Ye pump, and conduct a test about its fatigue resistance to observe impulse performance of the pump. Methods Luo- Ye pump is composed of five parts: pump body, diaphragm, diaphragm fixed ring, inlet and outlet of prosthetic valve. The injection forming technique is adopted to manufacture Luo-Ye pump body of auxiliary pump, and diaphragm fixed ring; the methods of cold forming and thermal fixing are adopted to make diaphragm of pump; inlet and outlet of pump are made by medical Tc-4 titanium; valves adopt 17# CL-I(standard form) domestic, artificial, mechanical valves. A Luo-Ye pump assembled undertakes fatigue test on analogue circulatory device. Results Pediatric Luo-Ye pump of 20 mL is manufactured by taking the method of injection moulding, in which the surface is smooth. The pump body is totally transparent. It is tested on analogue circulatory device. And no deformities and leakage are found around the pump body, diaphragm, inlet and outlet. Conclusion Adoption of injection moulding technique could successfully manufacture pediatric Luo-Ye pump, which has stable quality and realizes batch production. Pediatric Luo-Ye pump has satisfactory durability.
出处 《中国医疗设备》 2012年第10期56-59,共4页 China Medical Devices
基金 广东省教育部产学研结合基金资助项目(2008B090500210) "十二五"国家科技支撑计划课题(2011BAI11B22)
关键词 小儿罗叶泵 心室辅助装置 模拟循环台 心功能衰竭 pediatric Luo-Ye pump ventricular assist device analogue circulatory device cardiac failure
  • 相关文献

参考文献11

二级参考文献27

  • 1Chen CZ,Ye CX.Study on mechanical circulatory support and currently available supporting parts in China.Artif Organ,2006,30:510-513.
  • 2Ye CX,Umezu M,Nugent AH,et al.Invitro and invivo evaluation of the antithrombogenic properties of the spiral vortex pump for temporary left heart bypass.Perfusion,1991,6:123-129.
  • 3Ye CX,Umezu M,Nugent AH,et al.The spiral vortex pump:Design concept,flow pattern characteristic and results of invitro tests.Artif Organs.1990,14:46-49.
  • 4Song X,Wood HG,Day SW,et al.Studies of turbulence models in a computational fluid dynamics model of a blood pump.Artif Organs,2003,27:935-937.
  • 5Yamane Y,Miyamoto Y,Taijima K,et al.A comparative study between flow visualization and computational fluid dynamic analysis for the Sun Medical centrifugal blood pump.Artif Organs,2004,28:458-466.
  • 6Medvitz RB,Kreider JW,Manning KB,et al.Development and validation of a computational fluid dynamics methodology for simulation of pulsatile left ventricular assist devices.ASAIO J,2007,53:122-131.
  • 7Xiao XJ,Luo ZX,Ye CX,et al.The short-term pulsatile ventricular assist device for postcardiotomy cardiogenic shock:a clinical trial in China.Artif Organs,2009,33:373-377.
  • 8Hetzer R,Potapov EV,Stiller B,et al.Improvement in survival after mechanical circulatory support with pneumatic pulsatile ventricular assist devices in pediatric patients.Ann Thorae Surg,2006,82:917-925.
  • 9Kaczmarek I,Sachweh J,Groetzner J,et al.Mechanical circulatory support in pediatric patients with the MEDOS assist device.ASAIO J,2005,51:498-500.
  • 10Duncan BW.Pediatric mechanical circulatory support in the United States:past,present,and future.ASAIO J,2006,52:525-529.

共引文献25

同被引文献28

  • 1Pirouz S,Matthew H,Rossano JW, et al.Characteristics and outcomes of heart failure-related intensive care unit admissions in children with cardiomyopathy[J] .J Card Fail,2013,19(10):672- 677.
  • 2Rossano JW, Kim JJ,Decker JA,et al.Prevalence, morbidity, and mortality of heart failure-related hospitalizations in children in the United States:a population-based study[J].J Card Fail,2012, 18(6):459-470.
  • 3Zhu DM,Wang W, Chen H,et al.Left ventricular assist device for pediatric postcardiotomy cardiac failure[J].ASAlO J,2006, 52(5):603-604.
  • 4Mascio CE.The use of ventricular assist device support in children: the state of the art[J].ArtifOrgans,2015,39(1):14-20.
  • 5DeBakey ME.Lett ventricular bypass pump for cardiac assistance: Clinical experience[J].Am J Cardiol, 1971,27(1):3-11.
  • 6Blume ED,Naftel DC,Bastardi HJ,et al.Outcomes of children bridged to heart transplantation with ventricular assist devices:A multi-institutional study[J].Circulation,2006,113(19):2313-2319.
  • 7Husain SA,Wallis G,Fricker FJ,et al.Ventricular assist device implantation in the pediatric population:does pump size selection and associated hemodynamics impact outcomes[J]? ASAIO J,2008,54(5):474-478.
  • 8Luo JM,Chou NK,Chi NH,et al.Pediatric heart transplantation bridged with ventricular assist devices[J].Transplant Proc,2010, 42(3):913-915.
  • 9Lorts A,Zafar F, Adachi I,et al.Mechanical assist devices in neonates and infants[J].Semin Thorae Cardiovase Surg Pediatr Card Surg Annu,2014,17(1):91-95.
  • 10Kirklin JK, Bennett Pearce F, Dabal RJ,et al.Mechanical circulatory support:strategies and outcomes in pediatric congenital heart disease[J].Semin Thorac Cardiovase Surg Pediatr Card Surg Annu,2014,17(1 ):62-68.

引证文献1

二级引证文献6

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部