期刊文献+

离子色谱-抑制电导检测法测定环磷酰胺相关物质A的含量 被引量:4

Determination of Cyclophosphamide Related Substance A by Ion Chromatography-Suppressed Conductivity Detection
原文传递
导出
摘要 目的采用离子色谱.抑制电导检测法测定环磷酰胺相关物质A的含量。方法采用色谱柱lonPac CS12A(250mm×4.0mm Analytical),以0.16%甲烷磺酸为流动相,流速为1.0mL·min^-1,检测器为电导检测器,柱温为25℃,进样量为25μL。结果环磷酰胺相关物质A的定量下限为1.8μg·mL^-1,检测限为0.54μg·mL^-1;在1.8~24μg·mL^-1内线性关系良好(r〉0.999),在不同浓度水平范围内的平均回收率为97.45%~99.81%,RSD〈3%。结论本方法选择性好,灵敏度高,结果可靠,可满足环磷酰胺相关物质A检验分析的需要。 OBJECTIVE To establish ion exchange chromatography-suppressed conductivity detection for the determination of related substance A in cyclophosphamide. METHODS Using a Ionpac CS12A column (250 mm×4.0 mm Analytical), the mobile phase was 0.16% methane sulfonic acid, the flow rate was 1.0 mL·min^-1 with suppressed conductivity detector; the column temperature was 25 ℃; and the inject volume was 25 μL. RESULTS The detection and quantification limit was 0.54 and 1.8μg·mL^-1, respectively. The linear range for the content of degradation impurity A was 1.8-24 μg·mL^-1 (r〉0.999), the average recoveries were in the range of 97.45%-99.81% with the relative standard deviation below 3%. CONCLUSION The method is simple, rapid, accurate and is suitable for the determination of cyclophosphamide degradation impurity A.
出处 《中国现代应用药学》 CAS CSCD 2014年第5期567-570,共4页 Chinese Journal of Modern Applied Pharmacy
关键词 离子色谱法 抑制电导检测法 环磷酰胺 相关物质A ion chromatography suppressed conductivity detection cyclophosphamide related substance A
  • 相关文献

参考文献9

二级参考文献87

共引文献417

同被引文献67

引证文献4

二级引证文献20

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部