摘要
目的:评价与比较320排容积CT的前瞻性心电门控技术扫描和ECG管电流调制扫描技术的冠状动脉成像图像质量以及辐射剂量。方法:对临床拟诊为冠心病的97例患者,随机分成两组:①常规组(A组)53例,采用ECG管电流调制扫描技术;②低剂量组(B组)44例,采用前瞻性心电门控技术扫描。采用4分法主观评价冠状动脉可评估率,通过对比两组的升主动脉根部的平均强化CT值(SI)、噪声(SD)、信噪比(SNR)、对比信噪比(CNR)客观评价冠状动脉。采用Mann-Whitney U检验比较两组冠状动脉可评估率,采用两独立样本t检验比较两组的辐射剂量。结果:A组平均有效辐射剂量为(8.66±3.67)mSv,B组平均有效辐射剂量为(3.99±2.42)mSv,两组比较差异有统计学意义(t=-7.503,P<0.05)。A、B两组共1 187段血管,A组中较为优秀(3分或4分)的血管段数占92.4%(599/648),B组占92.9%(501/539),A、B两组冠状动脉图像质量主观评分比较差异无统计学意义(Z=-0.480,P=0.631)。A、B两组的SI、SD、SNR、CNR分别为(433.4±82.5)HU、22.1±6.2、20.7±5.7、16.6±5.0和(436.1±94.7)HU、21.5±4.9、20.9±5.5、16.9±4.7,两组冠状动脉段图像质量从客观上比较差异无统计学意义(t=0.153、-0.522、0.240、0.311,均P>0.05)。结论:320排容积CT冠状动脉成像采用低剂量技术可以减少54%的辐射剂量,又可保证图像质量。
tomography angiography (CCTA) received by prospective ECG-gated CTA mode with tube current modulated CFA mode on 320-slice volume CT.Methods: Ninety-seven patients with suspected coronary artery disease were randomly divided into two groups: 53 consecutive patients were performed CCTA with tube current modulated CFA mode (group A) and 44 patients with prospective electrocardiographic gating (group B). Evaluating coro-nary artery by subjective and objective judgements, the image quality of CCTA were subjectively evaluated using a 4-point scale, while objectively evaluated by comparing mean intraluminal attenuation (SI), noise (SD), signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR) of the ascending aorta. The image quality of coronary artery between groups was compared by Mann-WhitneyU test, and the radiation dose between groups was compared byt test.Results: The average radiation dose was (8.66±3.67) mSv in group A and (3.99±2.42) mSv in group B, with signiifcant difference for A vs B (t=-7.503,P〈0.05). Of 1187 coronary artery segments in group A and B, good image quality (score of 3 or 4) was achieved in 92.4% (599∕648) in group A, 92.9% (501/539) in group B, with no signiifcant difference for A vs B (Z=-0.480,P=0.631). The values of SI, SD, SNR, CNR in group A were (433.4±82.5)HU, (22.1±6.2), (20.7±5.7), (16.6±5.0), and the values of SI, SD, SNR, CNR in group B were (436.1±94.7) HU, (21.5±4.9), (20.9±5.5), (16.9±4.7). There was no signiifcant difference of the image quality between group A and B objectively (P〉0.05).Conclusion: 54% radiation dose of CCTA is reduced in 320-slice volume CT scan by using prospective ECG-gated CTA mode, moreover, the diagnostic quality of the image is also ensured.
出处
《温州医学院学报》
CAS
2014年第5期352-356,共5页
Journal of Wenzhou Medical College