期刊文献+

CT增强中对比剂注射方案探讨 被引量:4

A Research about Injection Project of Contrast Media During Enhanced CT Scanning
在线阅读 下载PDF
导出
摘要 目的 探讨CT增强中对比剂的最少使用量 ,且能保证高质量图像的方案。方法 对 5例肝门动态扫描CT增强图像有关靶部位的CT值逐层测量 ,从而得到相应的“CT值 时间曲线” ,最终找出有效增强周期。结果 根据经验确定主动脉、门静脉及肝实质的增强阈值为 70HU ,并找到增强阈值达到和结束的时间。结论 使用新型螺旋CT进行增强扫描 ,可以减少对比剂的使用量 ,且注射速率、延迟时间等有关参数也应根据具体情况确定。 Objective To investigate a method that perfect imaging could be obtained and minimal contrast media was used. Methods The 5 cases with enhanced CT images were measured at each slice, then the Hu of aim position was obtained and the Hu time curve was got. At last, the valid enhanced period would be found. Results According to the experience, the enhancement threshold of aorta, portal vein and hepar were set at 70 HU. The onset and ending time of enhancement threshold were found. Conclusion The dose of contrast media should be extenuatory when new type spiral CT was operated, and the correlative parameters, such as flow rate, delay time etc. should be adjusted according to the objective condition. [
出处 《中国医学影像技术》 CSCD 2001年第5期483-485,共3页 Chinese Journal of Medical Imaging Technology
关键词 CT增强 对比剂 注射量 Enhanced CT scanning Contrast media Volume
  • 相关文献

参考文献4

二级参考文献12

  • 1夏成德,郭俊渊,胡国栋.双期动态CT对肝内占位性病变诊断的研究[J].临床放射学杂志,1994,13(2):102-105. 被引量:8
  • 2孙振峰,孙钢.肝脏肿瘤的动态CT扫描[J].中华放射学杂志,1989,23(5):280-284. 被引量:4
  • 3林贵.肝肿瘤的微血管结构和血供[J].中华放射学杂志,1985,19(5):257-257.
  • 4曾斌,临床放射学杂志,1990年,9卷,234页
  • 5王仪生,中华放射学杂志,1990年,24卷,266页
  • 6孙振峰,中华放射学杂志,1989年,23卷,280页
  • 7林贵,中华放射学杂志,1985年,19卷,257页
  • 8Hwang G J,Radiology,1997年,202卷,383页
  • 9夏成德,临床放射学杂志,1994年,13卷,102页
  • 10王滨,临床放射学杂志,1990年,9卷,182页

共引文献25

同被引文献19

  • 1陈碧敏,李恒国.颈部肿块的影像学研究进展(讲座)[J].暨南大学学报(自然科学与医学版),2004,25(4):494-497. 被引量:4
  • 2丁兰芳,唐伟英.影响CT增强扫描效果的三要素[J].中国航天工业医药,1999,1(1):68-69. 被引量:2
  • 3窦新民,张建伟.双层螺旋CT颈部增强最佳注射扫描方案的研究[J].中国煤炭工业医学杂志,2005,8(9):1005-1006. 被引量:1
  • 4严福华,曾蒙,苏周,康荣.螺旋CT肝脏检查中技术参数的选择──时间-密度曲线的分析[J].临床放射学杂志,1996,15(5):290-293. 被引量:62
  • 5Computed Tomography Angiography (CTA),Radiologyinfo, www.radiologyinfo.org, 2003.7
  • 6Amstutz DR, Pitt AM, Partovi S, et al. Utility of A Dual Power Injector for Intracranial Computed Tomographic Angiography(CTA).Medrad CT Clinical Information, www. medrad. com/pdf/cranialcta.abstract.pdf ,2003
  • 7Mary Beth Massat, Contrast Media Injectors Growth of MRI and CT Procedures Drives Market Demand for Power Injectors,Imaging Technology News, www.reillycomm.com/it_archive/it_to0303_2.htm2003(3/4)
  • 8Injectors, Contrast Media, Angiography; Computed Tomography;Magnetic Resonance Imaging: ECRI HPCS; Jan. 2003
  • 9Hopper KD, Mosher TJ, Kasales C J, et al. Thoracic spiral CT: de livery of contrast material pushed with injectable saline solution in a power injector. Radiology, 1997,205:269-271
  • 10Haage P, Schmitz-Rode T, Hubner D, et al. Reduction of contrast material dose and artifacts by saline flush using a double power injector in helical CT of the thorax. AJR, 2000,164:1049-1053

引证文献4

二级引证文献6

相关作者