摘要
针对测量距离对辐射温度计检定结果的影响进行了实验研究,给出了几种所选择的常用的辐射温度计的实验数据和分析图表。确定了辐射温度计的最佳测量距离是在离开黑体辐射源靶面0.3~0.5 m范围内。
Aiming at the effect of measuring distance on calibration results of radiation thermometers, the experimental research was carried out. Both testing data and analytical charts for the several conventional radiation thermometers chosen are shown. The optimal measuring distance of radiation thermometer is defined, which is in the range of 0. 3 - 0. 5 m to be away from the target surface of blackbody radiation source.
出处
《计量学报》
CSCD
北大核心
2014年第6期588-590,共3页
Acta Metrologica Sinica
关键词
计量学
辐射温度计
测量距离
黑体辐射源
Metrology
Radiation thermometer
Measuring distance
Blackbody radiation source