期刊文献+

压力蒸汽灭菌3种B-D监测装置应用对比

Comparison of three B-D test devices in pressure steam sterilization
原文传递
导出
摘要 目的探讨目前医院在用的自制棉布标准B-D包、一次性B-D测试包和模拟管腔型监测装置3种压力蒸汽灭菌器监测装置的实用性能。方法严格按照ISO 11140-4/5:2006.12.14的测试方法,采用自制棉布标准B-D包、一次性B-D测试包和模拟管腔型监测装置,分别对压力蒸汽灭菌器进行监测,比较3种监测装置的监测结果、效率、成本等。结果一次性B-D测试包监测合格率最高(99.2%),平均每次耗费时间最短(2 min),管腔型监测装置平均每次使用费用最低(46.9元)。结论一次性B-D测试包和管腔型监测装置根据各自特点都是值得推荐的压力蒸汽灭菌器性能监测装置,而自制棉布标准B-D包由于敏感性、工作效率和成本各方面的原因,应当尽快升级换代。 Objective To compare the practicality performances of the standard cotton B- D test packs,diposable B- D test packs and Helix test products in hospital at present. Methods The standard cotton B- D test pack,diposable B- D test packs and Helix test products were used to test the pressure steam sterilization device according to ISO 11140- 4 /5: 2006. 12. 14 strictly,and the sensitivity,efficiency and cost of the three products were compared. Results Diposable B- D test packs got the highest qualified rate( 99. 2%) and took the shortest time( 2 minutes),while Helix test products cost the least( 46. 9 $) in all of the three products. Conclusion Diposable B- D test packs and Helix test products are two kinds of worth recommending devices in pressure steam sterilization,and because of its sensitivity,efficiency and cost,the standard cotton B- D test packs should be upgraded as soon as possible.
出处 《中国卫生检验杂志》 CAS 2016年第18期2677-2678,共2页 Chinese Journal of Health Laboratory Technology
基金 宁夏护理学会科研课题计划基金项目(NXHL15-20)
关键词 压力蒸汽灭菌 自制棉布标准B-D包 一次性B-D测试包 模拟管腔监测装置 Pressure steam sterilization Standard cotton B-D test packs Diposable B-D test packs Helix test products
  • 相关文献

参考文献10

二级参考文献27

  • 1张秀月,董丽华.B-D试验中常见技术差错及防范[J].中华医院感染学杂志,2004,14(7):790-790. 被引量:13
  • 2符爱娟,陈桂.B-D试验不合格原因分析[J].中国消毒学杂志,2005,22(1):112-113. 被引量:4
  • 3李斌,魏静蓉,王义辉,唐伟.脉动真空灭菌器B-D试验阳性结果的分析与验证[J].医疗卫生装备,2007,28(3):61-62. 被引量:3
  • 4AAMI/ISO 11140-05:2006-12-14,Sterilization of health care products-Chemical indicators-Part 5: Class 2 indicators for Bowie and Dick test sheets and packs.
  • 5EN285.Sterlization-Steam sterilizers-Large sterlizers.2006-05.
  • 6Bowie JH,Dick J.The Lancet, March 16,1963, 586-587.
  • 7Stuart J.Line and J.K.Pickerill,J.Clin.Path., 1973,26,716-720.
  • 8AAMI/ISO 11140-04:2006-12-14,Sterilization of health care products-Chemical indicators-Part 4: Class 2 indicators as an alternative to Bowie-Dick test fjor detection of steam penetration.
  • 9.消毒技术规范.[S].中华人民共和国卫生部,2002..
  • 10.消毒技术规范[S].[S].中华人民共和国卫生部,2002,11..