期刊文献+

空间磁偏转质谱计校准装置设计及校准方法研究 被引量:4

Design and calibration method for magnetic sector mass spectrometer calibration apparatus
在线阅读 下载PDF
导出
摘要 为确保磁偏转质谱计空间探测应用测量数据的准确性,设计了一套校准装置对磁偏转质谱计进行地面校准。本文介绍了装置的结构设计,并根据校准装置和磁偏转质谱计自身的特点,提出了二种校准方法,分别为:分压力测量法和标样气体法。研究表明,校准装置的极限真空度可达到5×10^(-8)Pa,二种校准方法对应不同的校准范围,可将质谱计的校准下限延伸至10^(-8)Pa,比国内目前已报道的质谱计的校准下限低1~2个量级。 In order to insure the measurement veracity of the space magnetic sector mass spectrometer, a calibration apparatus was designed to calibrate the mass spectrometer on the ground. In this paper, the calibration apparatus structure design was given.Additionally, two calibration methods were proposed according to the magnetic sector mass spectrometer characteristics. The two methods were the standard gas calibration method and the partial pressure measurement method, respectively. The preliminary results show that ultimate pressure 10^(-8)Pa can be achieved by the apparatus. The two calibration methods have different calibration range, and the calibration limit can be extended to 5×10^(-8)Pa, which is lower one or two orders of magnitude than that reported by the literatures.
作者 董猛 成永军 李得天 孙雯君 郭美如 赵澜 张瑞芳 吉康 孙健 DONG Meng CHENG Yong-jun LI De-tian SUN Wen-jun GUO Mei-ru ZHAO Lan ZHANG Rui-fang JI Kang SUN Jian(Lanzhou Institute of Physics, Lanzho)
出处 《真空》 CAS 2016年第6期28-33,共6页 Vacuum
关键词 磁偏转质谱计 空间探测 校准装置 校准方法 Magnetic sector mass spectrometer Space exploration Calibration apparatus Calibration method
  • 相关文献

参考文献5

二级参考文献31

  • 1查良镇,朱秀珍,陈旭,余海棠.四极质谱计校准和应用研究系统[J].真空科学与技术,1993,13(6):405-412. 被引量:17
  • 2陈旭,张启亮,谷京华,朱秀珍,查良镇.一种新型四极质谱计及其性能测试[J].真空,1995,32(5):16-23. 被引量:6
  • 3Basford J A, Boeckmann M D, Ellefson R E. American vacuum society recommended practice: Recommended practice for the calibration of mass spectrometers for partial pressure analysis[J]. J. Vac. Sci. Technol. , 1993,A11(3):A22.
  • 4Mattinson, J. M. , 1994. A study of complex discordance in zircons using step-wise dissolution techniques. Contrib.Mineral. Petrol. , 116:117- 129.
  • 5Mezger, K. , Krogstad, E. J. , 1997. Interpretation of discordant U-Pb zircon ages:An evaluation. J. Metamorph. Geol. ,15:127-140.
  • 6Müller,W. , Mancktelow, N. S. , Meier, M. , 2000. Rb-Sr microchrons of synkinematic mica in mylonites:An example from the DAV fault of the Eastern Alps. Earth Planet. Sci. Lett. , 180: 385-397.
  • 7Müiller,W., 2003. Strengthening the link between geochronology, textures and petrology. Earth Planet. Sci.Lett. ,206:237-251.
  • 8Poller, U. , Liebetrau, V. , Todt, W. , 1997. U-Pb singlezircon dating under cathodoluminescence control (CLC-method) :Application to polymetamorphic orthogneisses. Chem. Geol. ,139:287-297.
  • 9Sergeev,S. A. ,Komarov, A. N. , Bickel, R. A. , et al., 1997.A new microtome for cutting hard submillimeter-sized crystalline objects for promoting high-resolution instrumental microanalysis. Eur. J. Mineral. , 9: 449-456.
  • 10Sharma, M. , Papanastassiou, D. A. , Wasserburg, G. J. , et al. ,1996. The issue of the terrestrial record of 146Sm.Geochim. Cosmochim. Acta, 60: 2037- 2047.

共引文献66

同被引文献173

引证文献4

二级引证文献19

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部