期刊文献+

批量测定某矿区尾矿中铜、锌、锡方法比对

Comparison of methods for the determination of copper,zinc and tin in tailings of a mining area
在线阅读 下载PDF
导出
摘要 本文分别建立了X射线荧光光谱法与电感耦合等离子体发射光谱法同时测定某矿区尾矿中铜、锌、锡三种元素。对比两种方法的精密度,准确度,检出限、检测耗时。检测结果表明,两种方法的相对误差均小于5%,相对标准偏差均小于3%,准确度和精密度均较好。X射线荧光光谱法的检出限更低。电感耦合等离子体发射光谱法可以实现批量处理样品、检测样品,耗时短,效率高,具有优势。 This paper introduces two methods about X-ray fluorescence spectrometry and inductively coupled plasma atomic emission spectrometry for the simultaneous determination of copper,zinc and tin in tailings of a mining area.The precision,accuracy,detection limit and detection time of the two methods were compared.The results show that the relative errors of the two methods are less than 5%and the relative standard deviation are less than 3%.The accuracy and precision are good.The detection limit of X-ray fluorescence spectrometry is low.Inductively coupled plasma atomic emission spectrometry can process and detect samples in batches,which is time-consuming and efficient.Therefore,inductively coupled plasma emission spectrometry has more detection advantages.
作者 董艳敏 任丽芳 杨银柱 孙铭旋 张金磊 杨莹芳 DONG Yan-min;REN Li-fang;YANG Yin-zhu;SUN Ming-xuan;ZHANG Jin-lei;YANG Ying-fang(Shandong Gold Geologyand Mineral Exploration Linited Corporation,Shandong Laizhou 261400,China)
出处 《世界有色金属》 2021年第18期228-230,共3页 World Nonferrous Metals
关键词 X射线荧光光谱法 电感耦合等离子体发射光谱法 尾矿 方法对比 X-ray fluorescence spectrometry inductively coupled plasma atomic emission spectrometry tailings method comparison
  • 相关文献

参考文献9

二级参考文献70

共引文献109

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部