小野对动态调强放疗计划剂量分布的影响

Effect of Small Field on Dose Distribution in Dynamic Intensity Modulated Radiotherapy Plan
在线阅读 下载PDF
导出
摘要 目的探讨小野对动态调强放疗计划剂量分布的影响。方法随机选取2020年医院放疗技术科的20例肿瘤患者的动态调强放疗计划,分别采用含实测小野数据的AAA算法和不含实测小野数据的AAA算法计算其剂量分布,比较两种算法的通过率。结果以3 mm/3%为标准,含实测小野数据的AAA算法通过率均高于不含实测小野数据的AAA算法,对两种算法的计划验证结果进行配对t检验,差异有统计学意义(P<0.05)。结论在动态调强放疗计划中多叶准直器连续运动所形成的小野对剂量分布产生了影响,有条件的情况下建议实测加速器的小野剂量学数据。 Objective The effect of small field on dose distribution in dynamic intensity modulated radiotherapy plan was investigated. Methods The dynamic intensity modulated radiotherapy plans of 20 patients in the Department of Radiotherapy Technology of the hospital in 2020 was randomly selected, the AAA algorithms with and without measured data of small field were both used to calculate the dose distribution, and the pass rates of the two algorithms were compared.Results The pass rate of the AAA algorithm with measured data of small field were all higher than those of the AAA algorithm without measured data of small field with the standard of 3 mm/3%. There was a significant difference between the two algorithms in the results of paired t-test(P<0.05). Conclusion The small fields formed by the continuous movement of the multi-leaf collimator in the dynamic intensity modulated radiotherapy plan have an effect on the dose distribution. If possible, it is recommended to measure the small field dosimetry data of the accelerator.
作者 袁清伟 吴瑶 陈涛 郝连花 林清森 赵汝彬 Yuan Qingwei;Wu Yao;Chen Tao;Hao Lianhua;Lin Qingsen;Zhao Rubin(Department of Radiotherapy Technology,Linyi People's Hospital,Shandong Province,Linyi Shandong 276000,China)
出处 《医疗装备》 2021年第23期6-8,共3页 Medical Equipment
关键词 小野 动态调强放疗计划 剂量分布 Small field Dynamic intensity modulated radiotherapy plan Dose distribution
  • 相关文献

参考文献3

二级参考文献13

  • 1Chen SW, Wang XT, Chen LX, et al. Monte Carlo evaluations of the absorbed dose and quality dependence of AL203 in radiotherapy photon beams. Med Phys,2009,36:4421-4424.
  • 2Zhu XR, Allen JJ, Shi J, et al. Total scatter factors and tissue maximum ratios for small radiosurgery fields : comparison of diode detectors, a parallel-plate ion chamber, and radiographic film. Med Phys ,2000,27:472-477.
  • 3McKerracher C, Thwaites DI. Assessment of new small-field detectors against standard-field detecors for practical stereotactic beam data acquisition. Phys Med Biol, 1999,44:2143-2160.
  • 4Cheng CW , Cho SH,Taylor M, et al. Determination of zero-field size percent depth doses and tissue maximum ratios for stereotactic radiosurgery and IMRT dosimetry : comparison between experimental measurement and Monte Carlo simulation. Med Phys, 2007,34:3149-3157.
  • 5Zhu XR, Allen JJ, Shi J, et al. Total scatter factors and tissue maximum ratios for small radiosurgery fields : comparison of diode detectors, a parallel-plate ion chamber, and radiographic film. Med Phys ,2000,27:472-477.
  • 6Verhaegen F, Das IJ, Palmans H. Monte Carlo dosimetry study of 6 MV stereotactic radiosurgery unit. Phys Med Biol, 1998,43: 2755 -2768.
  • 7张晓军,王建华,涂彧.二维电离室矩阵在调强适形放疗剂量学验证中的应用[J].国际放射医学核医学杂志,2007,31(5):318-320. 被引量:13
  • 8李智华,张瑞平,邓满国,蒋胜鹏.医科达多叶准直器到位精度的胶片测量方法[J].中华放射肿瘤学杂志,2008,17(1):66-68. 被引量:7
  • 9王运来,戴相昆,谢耩.二维电离室矩阵在调强放疗剂量验证中的应用[J].中华放射肿瘤学杂志,2008,17(2):150-151. 被引量:19
  • 10蒋胜鹏,李智华.加速器机架角度对多叶准直器叶片到位精度的影响[J].中华放射肿瘤学杂志,2009,18(4):317-320. 被引量:15