一种用于验证胸腔LDCT-PCNB扫描参数模型的设计与应用

Design and Application of a Parameter Model for Verification of Thoracic LDCT-PCNB Scanning
在线阅读 下载PDF
导出
摘要 目的在诸多涉及胸腔低剂量CT引导下经皮穿刺活检(Low-Dose CT-Guided Percutaneous Core Needle Biopsy,LDCT-PCNB)扫描参数的研究里,其设计多为经验法或者穷举法,考虑到不同机型硬件差异,这些研究的通用性及伦理有待商榷。为此设计一款专用于验证胸部LDCT-PCNB扫描参数的体模。方法根据胸腔天然对比以及CT线性衰减原理,通过SKUP(SketchUp草图大师)设计模型,并基于熔融沉积制造(Fused Deposition Modeling,FDM)的3D打印机来制作实体,带入CT进行参数对比试验,对不同低剂量参数结果进行t检验,选择数值较好LDCT参数带入实际扫描,并使用t检验与普通扫描对比。结果经过体模对比试验,两种LDCT参数结果差异较大(P>0.01),而在实际穿刺中,体模实验选择的参数比普通扫描其ED降低了约25倍,且两者质量剂量比差异较大(P=0.01)。结论该体模可一定程度模拟胸腔穿刺的几何关系及相关参数,并择优选择质量剂量比较高的参数用于实际使用,有效减少了重复实验,提高辐射使用效率,并避免伦理问题。 Objective In many studies on scanning parameters of low-dose CT-guided percutaneous core needle biopsy(LDCT-PCNB),the design was mostly empirical method or exhaustive method.considering the difference of hardware of different models,the generality and ethics of these studies were to be discussed.Therefore,a phantom was designed to verify the scanning parameters of LDCT-PCNB.Methods According to the principle of chest natural contrast and CT linear attenuation,a model was designed by SKUP,and the 3D printer based on fused deposition modeling(FDM)was used to make the entity,which was brought into CT for parameter contrast test.The results of different low-dose parameters were tested by t-test,and the low-dose parameters with better values were selected to be brought into the actual scanning,and compared with the ordinary scanning by t-test.Results Through the phantom comparison test,there were significant differences between the two LDCT parameters(P>0.01).In the actual puncture,the ED parameters selected in the phantom experiment were about 25 times lower than those in the conventional scanning,and the difference of the mass dose ratio between the two was large(P=0.01).Conclusion The phantom can simulate the geometric relationship and related parameters of thoracic puncture to a certain extent,and select parameters with high quality and dose for practical use,which can effectively reduce repeated experiments,improve the efficiency of radiation use,and avoid ethical problems.
作者 赵子健 李若梅 吴瑞祥 叶亮 沈雪芹 陆璐 魏夏红 ZHAO Zijian;LI Ruomei;WU Ruixiang;YE Liang;SHEN Xueqin;LU Lu;WEI Xiahong(Department of Imaging,Hefei Hospital Affiliated to Anhui Medical University,Hefei Anhui 230011,China;Department of General,Hefei Ion Medical Center,University of Science and Technology of China,Hefei Anhui 230011,China)
出处 《中国医疗设备》 2022年第3期52-55,70,共5页 China Medical Devices
基金 安徽医科大学附属合肥医院(市二院)科研基金(2019-46) 合肥市二院光华青年科研基金(2020-178.28)。
关键词 低剂量CT引导下经皮穿刺活检 熔融沉积制造 CT引导穿刺 质量剂量比 3D打印 low-dose CT-guided percutaneous core needle biopsy fused deposition modeling CT guided puncture mass dose ratio 3D printing
  • 相关文献

参考文献16

二级参考文献131

共引文献153

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部