鼻黏膜纤毛显微内窥镜光学系统设计 被引量:3

Design of Optical System for Nasal Mucociliary Microendoscope
原文传递
导出
摘要 为了实现从前鼻孔进镜对鼻黏膜表面进行直接显微成像,从而进行无创性观察和测量在体鼻纤毛运动,设计了一种能够分辨鼻纤毛,并且带有30°视向角的变倍硬性显微内窥镜。所设计的硬性内窥镜系统物方数值孔径为0.15,分辨率为272 lp/mm,工作距离为3.00 mm,视向角为30°,物面高度为0.40 mm,系统总长度为205.62 mm,物方口径为4.65 mm,放大倍率为6×~10×。此鼻黏膜纤毛显微内窥镜将可以避免取材造成的纤毛功能损害和受试者痛苦,极大地提高纤毛相关疾病的临床诊断能力,成为助益纤毛领域科研和临床工作的重要突破。 Objective Motile cilia of nasal mucosa are widely distributed on the mucosal surface of the human respiratory tract.The defense function of removing mucus and pathogenic particles from the mucosal surface can be realized by swinging regularly in a specific direction.The normal operation of this function is important for maintaining respiratory tract health and human health.The dysfunction of cilia will lead to a series of pathological manifestations,which will seriously affect the health and quality of life of patients.The traditional ciliary motion assessment methods are invasive and cannot reflect the actual motion state of the body.In order to perform direct microscopic imaging of the nasal mucosal surface through the anterior nostril for non-invasive observation and measurement of nasal ciliary movement in vivo,we design a variable-magnification rigid microendoscope with a viewing angle of 30°.The nasal mucosal ciliary microendoscope will avoid the damage to ciliary function and the pain of subjects caused by material extraction,which thus greatly improves the clinical diagnosis ability of cilia-related diseases and becomes an important breakthrough in the scientific research and clinical work in the field of cilia.Methods The design of the optical system of the nasal mucociliary microendoscope is mainly divided into three aspects:optical system design,image quality analysis,and tolerance analysis.The optical system mainly includes an objective lens system,relay system,eyepiece system,and variational adapter system.Firstly,a prism structure is determined so that it has a viewing angle of 30°.On the premise of ensuring the object resolution,we reduce the circumscribed circle diameter of the prism as much as possible and then reduce the object diameter of the entire endoscope.According to the structure of the cilia,we determine the numerical aperture of the objective lens system to provide a sufficient margin to avoid machining errors in the actual processing.According to the principle of pupil matching,the object image side,the object image side of the relay system,and the object side of the eyepiece all adopt a telecentric optical path.The relay system adopts the HOPKINS lens whose fully symmetrical structure can realize equal-proportion image transmission,and the vertical axis aberration can be automatically reduced.A piece of the negative lens can be separated to make it a thick lens,so as to solve the problem of excessive field curvature.The whole endoscope system has no visual requirement.An integrated design method of eyepiece and variable-magnification adapter systems is proposed,which can simplify the structure,reduce the cost,and better correct the aberration.In order to obtain a complete rigid endoscope optical system,it is necessary to connect all parts of the structure.Each part is connected in the order of objective lens system,relay system,eyepiece system,and variable-magnification adapter system.During the connection process,we make a simple optimization to ensure that the positions of each image plane are in the air.In addition,in order to avoid the loss of light energy,operands are still used to control the telecentricity of each image plane,and the system magnification of the connected endoscope is controlled by the image plane height.Finally,we further optimize the connected optical system and use operands to reduce field curvature and distortion,control glass thickness and air spacing,and make it machinable and assembling.MTF curve,spot diagram,and field curve distortion diagram are selected as the image quality evaluation criteria of the system to determine the optimized image quality.Tolerance analysis is carried out according to the given tolerance value to meet the performance requirements and minimize the production cost.Results and Discussions In this paper,we propose a method for direct observation of nasal mucosa cilia by a rigid microendoscope,which can avoid the damage of cilia functions and the pain of subjects caused by sampling and greatly improve the clinical diagnostic ability of cilia-related diseases.The numerical aperture of the system is 0.15,which greatly improves the resolution ability of the system.The relay system changes the HOPKINS structure to make a negative lens separated into a thick lens to effectively correct the field curvature.The system is designed as a variable-magnification structure,which can magnify objects as required.In this paper,the integrated design of the eyepiece system and the variable-magnification adapter system,as well as that of the whole system are proposed to simplify the structure,correct the aberration to the maximum extent,and achieve the imaging quality to the diffraction limit.Conclusions Aiming at the invasiveness and inaccuracy of the existing methods for evaluating nasal ciliary motion,we propose a method of direct microscopic imaging of the nasal mucosal surface through the anterior nostril for non-invasive observation and measurement of nasal ciliary movement in vivo.We use an integrated design method to design a microendoscope system with a viewing angle of 30°,high resolution,and variable magnification.The system achieves the goal of a viewing angle of 30°by secondary reflection on the viewing prism.The working wavelength is the visible light band.The working distance of the system is 3 mm,and the resolution is 272 lp/mm.The object's surface height is 0.4 mm,and the object's square aperture is 4.65 mm.In addition,the magnification is 6×-10×.We evaluate the image quality and find that the defocused spots of the three structures are smaller than the Airy disk,and the MTF curve reaches the diffraction limit.We analyze the tolerance,and the results show that it meets the processing conditions.The endoscope system is of great significance for non-invasive observation and study of nasal mucosal cilia in vivo.
作者 王鑫 向阳 冯大伟 杨凯明 逄川 陈雷 Wang Xin;Xiang Yang;Feng Dawei;Yang Kaiming;Pang Chuan;Chen Lei(School of Optoelectronic Engineering,Changchun University of Science and Technology,Changchun 130022,Jilin,China;Department of Otorhinolaryngology Head and Neck Surgery,Chinese PLA General Hospital,Beijing 100039,China;National Clinical Medical Research Center for Otorhinolaryngology,Beijing 100039,China)
出处 《光学学报》 EI CAS CSCD 北大核心 2023年第4期166-175,共10页 Acta Optica Sinica
关键词 医用光学 鼻黏膜纤毛 硬性内窥镜 视向角 变倍 高分辨率 medical optics nasal mucosa cilia rigid endoscope viewing angle zoom high resolution
  • 相关文献

参考文献5

二级参考文献30

  • 1石国华.内窥镜技术发展简述[J].医疗保健器具,2005(1):53-53. 被引量:3
  • 2贾晓航,颜青来,文燕.医用硬性内窥镜畸变的评定基础和方法[J].光学学报,2006,26(8):1226-1230. 被引量:17
  • 3白闻喜.医用内窥镜视放大率的测量原理及测试方法.长春光学精密机械学院学报,1985,8(3):84-88.
  • 4Zobel,Jurgen.Endoscope optical system United States[P].1988,4:723-843.
  • 5向阳.电视腹腔镜系统[D].长春:长春光学精密机械学院,1996.
  • 6向阳,姜会林,胡家升.目视医用硬性内窥镜光学系统初始结构设计[J].长春理工大学学报(自然科学版),2007,30(3):5-6. 被引量:10
  • 7T H Tomkinson, J L Bentley, M K Crawford, el al.. Rigid endoscopic relay systems: a comparative study [J]. Appl ()pt. 1996, 35(34): 6674--6683.
  • 8R Liang. Optical Design for Biomedical Imaging [ M]. New York: SPIEPress, 2010. 379--386.
  • 9F Doglietto, D M Prevedello, J A Jane Jr, et al.. A brief history of endoscopic transsphenoidal surgery: from Philipp Bozzini to the First World Congress of Endoscopic Skull Base Surgery [J]. Neurosurgical Focus, 2005, 19(6): 1--6.
  • 10J Bhatt, A Jones, S Foley, et a[.. Harold Horace Hopkins: a short biography [J]. BJU International, 2010, 106(10): 1425-- 1428.

共引文献35

同被引文献19

引证文献3

二级引证文献1

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部