颈椎间盘假体稳定性生物力学评价 被引量:10

Biomechanical evaluation of the stability of cervical intervertebral disc prosthesis
在线阅读 下载PDF
导出
摘要 目的借鉴腰椎人工椎间盘置换成熟的经验,评价颈椎人工椎间盘置换对失稳颈段脊柱稳定性的重建作用。方法在6具新鲜尸体的颈椎标本上切除C5~6椎间盘,将人工颈椎间盘假体植入该间隙内。对该颈椎节段进行前屈/后伸、左/右侧弯和左/右旋转运动稳定性评价,经颈椎三维运动测量系统测出该节段的运动范围(ROM)。随后施加2Nm,0.6Hz的疲劳载荷共2500次,评定其稳定性,并与标准的C5~6椎间盘破坏后稳定性值相比较。结果正常C5~6前屈、后伸运动范围平均值为17.03°;人工椎间盘植入后前屈、后伸运动范围7.80°,疲劳载荷后前屈、后伸运动范围9.98°;椎间盘破坏后前屈、后伸运动范围25.53°,人工椎间盘植入具有显著的稳定性。正常左右侧弯运动范围平均值为10.75°;人工椎间盘植入后和疲劳载荷后左右侧弯运动范围分别为8.43°和9.43°,椎间盘破坏后左、右侧弯运动范围20.70°。正常左、右旋运动范围平均值为14.82°;人工椎间盘植入后和疲劳载荷后左、右旋运动范围分别为13.38°和13.83°,椎间盘破坏后左右旋运动范围25.88°。结论颈椎间盘破坏后椎节稳定性显著性破坏。人工椎间盘植入后脊柱在前屈、后伸,左右侧弯方向上稳定性显著性改善,左右旋转接近正常值,并可提供59%正常屈伸活动,88%正常左右侧弯活动,93%正常左右旋? Aim To evaluate the stability of the cervical intervertebral disc prosthesis. Methods Six fresh frozen cadaveric C5- 6 motion segments with intact ligaments were used. The C? - 6 discs were resected and implanted with cervical intervertebral disc prosthesis. A cyclic load was applied to the specimens at the rate of 0.6 Hz up to 2 500 cyclis. The ranges of motion (ROMs) of the segments were measured using stereophotogrammetry before and after the prosthesis replacement. Results The normal mean ROM of C? 6 anteoflexion/post extension was 17.03° and 7.8° respectively before and after the prosthesis replacement. After fatigue load it was 9.98° , and after disc damage it was 25.53° ( P< 0.01) . The normal mean ROM of C5? 6 left/right lateral bending was 10.75° . The mean ROMs of C5- 6 left/right lateral bending were 8.43° , 9.43° and 20.70° after the prosthesis replacement, fatigue load and disc damage respectively( P< 0.01) . The normal mean ROM of C5? 6 left/right axial rotation was 14.82° . The mean ROMs of C5? 6 left/right axial rotation were 13.38° , 13.83° and 25.88° after the prosthesis replacement, fatigue load and disc damage( P< 0.01) . Conclusion After the disc destroyed, the stability of the motion segment decreased significantly. After cervical intervertebral disc prosthesis replacement, it is significantly improved in the stability of anteflexion/post extension and left/right lateral bending. The mean ROM of C5- 6 left/right axial rotation is close to the normal value. <P>
出处 《中国临床康复》 CSCD 2003年第17期2388-2389,共2页 Chinese Journal of Clinical Rehabilitation
基金 上海市科技发展基金资助项目(编号:024419075)~~
关键词 颈椎间盘假体 稳定性 生物力学 颈椎人工椎间盘置换 颈椎病 cervical vertebrae biomechanics cervical spondylosis bioprosthesis
  • 相关文献

参考文献7

  • 1Wigfield C, Gill S, Nelson R, Langdon I, Metcalf N, Robertson J. Influence of an artificial cervical joint compared with fusion on adjacent-level motion in the treatment of degenerative cervical disc disease. J Neurosurg 2002; 96(1 Suppl) : 17 - 21.
  • 2Katsuura A, Hukuda S, Saruhashi Y, Mori K. Kyphotic malalignment after anterior cervical fusion is one of the factors promoting the degenerative process in adjacent intervertebral levels. Eur Spine J 2001: 10(4): 320 -4.
  • 3Hochschuler SH, Ohnmeiss DD, Guyer RD, Blumenthai SL. Artificial disc: preliminary results of a prospective study in the United States. Eur Spine J 2002; 11 Suppl 2:S106 - 10.
  • 4White III AA, Panjabi MM. Clinical Biomechanics of the Spine. 2nd ed,Philadelphia: Lippincott Williams and Wilkins 1990:100 - 1.
  • 5Zhu QA. Three-dimensional motion of the middle and lower cervical spine.Zhongguo Linchuang Jiepouxue Zazhi( Chin J Clin Anat) 1992; 10(4)300 -3.
  • 6Golaab A, Birchall D, Laitt RD, Jackson A.Significance of CSF area measurements in ervical spondylitic myelopathy. Br J Neurosurg 2001; 15(1): 17 - 21.
  • 7Emery SE. Cervical spondylotic myelopathy: diagnosis and treatment. J Am Acad Orthop Surg 2001;9(6): 376 -88.

同被引文献110

引证文献10

二级引证文献38

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部